DC Lead-Off  
In this method, the lead-off excitation is with a dc signal. The dc excitation signal can be chosen from either an  
external pull-up or pull-down resistor or a current source or sink, as shown in Figure 50. One side of the channel  
is pulled to supply and the other side is pulled to ground. The internal current source and current sink can be  
swapped by setting the FLIP1 and FLIP2 bits in the LOFF_SENS register. In case of current source or sink, the  
magnitude of the current can be set by using the ILEAD_OFF[1:0] bits in the LOFF register. The current source  
or sink gives larger input impedance compared to the 10-Mpull-up or pull-down resistor.  
AVDD  
AVDD  
Device  
Device  
10 MW  
INP  
INN  
INP  
INN  
PGA  
PGA  
10 MW  
a) External Pull-Up/Pull-Down Resistors  
b) Input Current Source  
Figure 50. DC Lead-Off Excitation Options  
Sensing of the response can be done either by looking at the digital output code from the device or by monitoring  
the input voltages with an on-chip comparator. If either of the electrodes is off, the pull-up resistors and the pull-  
down resistors saturate the channel. By looking at the output code it can be determined that either the P-side or  
the N-side is off. To pinpoint which one is off, the comparators must be used. The input voltage is also monitored  
using a comparator and a 4-bit digital-to-analog converter (DAC) whose levels are set by the COMP_TH[2:0] bits  
in the LOFF register. The output of the comparators are stored in the LOFF_STAT register. These two registers  
are available as a part of the output data stream. (See the Data Output Protocol (DOUT) subsection of the SPI  
Interface section.) If dc lead-off is not used, the lead-off comparators can be powered down by setting the  
PD_LOFF_COMP bit in the CONFIG2 register.  
An example procedure to turn on dc lead-off is given in the Lead-Off subsection of the Quick-Start Guide section.  
AC Lead-Off  
In this method, an out-of-band ac signal is used for excitation. The ac signal is generated by alternatively  
providing an internal current source and current sink at the input with a fixed frequency. The excitation frequency  
is a function of the output data rate and is fDR/4. This out-of-band excitation signal is passed through the channel  
and measured at the output.  
Sensing of the ac signal is done by passing the signal through the channel to digitize it and measure at the  
output. The ac excitation signals are introduced at a frequency that is above the band of interest, generating an  
out-of-band differential signal that can be filtered out separately and processed. By measuring the magnitude of  
the excitation signal at the output spectrum, the lead-off status can be calculated. Therefore, the ac lead-off  
detection can be accomplished simultaneously with the ECG signal acquisition.  