07 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet
MPSMU Module Specifications
Voltage range, resolution, and accuracy (high resolution ADC)
Voltage
range
Force
resolution
Measure
resolution
Force accuracy 1
±(% + mV)
Measure accuracy 1
±(% + mV)
Maximum
current
±0.5 V
±2 V
±5 V
±20 V
±40 V
±100 V
25 μV
100 µV
250 μV
1 mV
2 mV
5 mV
0.5 μV
2 µV
5 μV
20 µV
40 µV
100 µV
±(0.018 + 0.15)
±(0.018 + 0.4)
±(0.018 + 0.75)
±(0.018 + 3)
±(0.018 + 6)
±(0.018 + 15)
±(0.01 + 0.12)
±(0.01 + 0.14)
±(0.009 + 0.25)
±(0.009 + 0.9)
±(0.01 + 1)
100 mA
100 mA
100 mA
100 mA
2
2
±(0.012 + 2.5)
1. ± (% of reading value + offset value in mV)
2. 100 mA (Vo ≤ 20 V), 50 mA (20 V < Vo ≤ 40 V), 20 mA (40 V < Vo ≤ 100 V), Vo is the output voltage in V.
Current range, resolution, and accuracy (high resolution ADC)
Current
range
Force
resolution
Measure
resolution
Force accuracy 1
±(% + A + A)
Measure accuracy 1
±(% + A + A)
Maximum
voltage
±1 nA
50 fA
500 fA
5 pA
50 pA
500 pA
5 nA
50 nA
500 nA
5 µA
10 fA
10 fA
±(0.1 + 3E-13 + Vo x 1E-15)
±(0.1 + 3E-12 + Vo x 1E-14)
±(0.05 + 3E-11 + Vo x 1E-13)
±(0.05 + 3E-10 + Vo x 1E-12)
±(0.05 + 3E-9 + Vo x 1E-11)
±(0.035 + 15E-9 + Vo x 1E-10)
±(0.04 + 15E-8 + Vo x 1E-9)
±(0.04 + 15E-7 + Vo x 1E-8)
±(0.045 + 15E-6 + Vo x 1E-7)
±(0.1 + 2E-13 + Vo x 1E-15)
±(0.1 + 1E-12 + Vo x 1E-14)
±(0.05 + 2E-11 + Vo x 1E-13)
±(0.05 + 1E-10 + Vo x 1E-12)
±(0.04 + 2E-9 + Vo x 1E-11)
±(0.03 + 3E-9 + Vo x 1E-10)
±(0.03 + 6E-8 + Vo x 1E-9)
±(0.03 + 2E-7 + Vo x 1E-8)
±(0.04 + 6E-6 + Vo x 1E-7)
100 V
100 V
100 V
100 V
100 V
100 V
100 V
±10 nA
±100 nA
±1 µA
±10 µA
±100 µA
±1 mA
100 fA
1 pA
10 pA
100 pA
1 nA
±10 mA
±100 mA
10 nA
100 nA
100 V
2
1. ± (% of reading value + fixed offset in A + proportional offset in A), Vo is the output voltage in V.)
2. 100 V (Io ≤ 20 mA), 40 V (20 mA < Io ≤ 50 mA), 20 V (50 mA < Io ≤ 100 mA), Io is the output current in A.
Voltage range, resolution, and accuracy (high speed ADC)
Voltage
range
Force
resolution
Measure
resolution
Force accuracy 1
±(% + mV)
Measure accuracy 1
±(% + mV)
Maximum
current
±0.5 V
±2 V
±5 V
±20 V
±40 V
±100 V
25 μV
100 µV
250 μV
1 mV
2 mV
5 mV
25 μV
100 µV
250 μV
1 mV
2 mV
5 mV
±(0.018 + 0.15)
±(0.018 + 0.4)
±(0.018 + 0.75)
±(0.018 + 3)
±(0.018 + 6)
±(0.018 + 15)
±(0.01 + 0.25)
±(0.01 + 0.7)
±(0.01 + 2)
±(0.01 + 4)
±(0.015 + 8)
±(0.02 + 20)
100 mA
100 mA
100 mA
100 mA
2
2
1. ±(% of reading value + offset value in mV). Averaging is 128 samples in 1 PLC.
2. 100 mA (Vo ≤ 20 V), 50 mA (20 V < Vo ≤ 40 V), 20 mA (40 V < Vo ≤ 100 V), Vo is the output voltage in V.
Current range, resolution, and accuracy (high speed ADC)
Current
range
Force
resolution
Measure
resolution
Force accuracy 1
±(% + A + A)
Measure accuracy 1
±(% + A + A)
Maximum
voltage
±1 nA
50 fA
500 fA
5 pA
50 pA
500 pA
5 nA
50 nA
500 nA
5 µA
50 fA
500 fA
5 pA
50 pA
500 pA
5 nA
50 nA
500 nA
5 µA
±(0.1 + 3E-13 + Vo x 1E-15)
±(0.1 + 3E-12 + Vo x 1E-14)
±(0.05 + 3E-11 + Vo x 1E-13)
±(0.05 + 3E-10 + Vo x 1E-12)
±(0.05 + 3E-9 + Vo x 1E-11)
±(0.035 + 15E-9 + Vo x 1E-10)
±(0.04 + 15E-8 + Vo x 1E-9)
±(0.04 + 15E-7 + Vo x 1E-8)
±(0.045 + 15E-6 + Vo x 1E-7)
±(0.25 + 3E-13 + Vo x 1E-15)
±(0.25 + 2E-12 + Vo x 1E-14)
±(0.1 + 2E-11 + Vo x 1E-13)
±(0.1 + 2E-10 + Vo x 1E-12)
±(0.05 + 2E-9 + Vo x 1E-11)
±(0.05 + 2E-8 + Vo x 1E-10)
±(0.04 + 2E-7 + Vo x 1E-9)
±(0.04 + 2E-6 + Vo x 1E-8)
±(0.1 + 2E-5 + Vo x 1E-7)
100 V
100 V
100 V
100 V
100 V
100 V
100 V
±10 nA
±100 nA
±1 µA
±10 µA
±100 µA
±1 mA
±10 mA
±100 mA
100 V
2
1. ±(% of reading value + fixed offset in A + proportional offset in A), Vo is the output voltage in V.)
2. 100 V (Io ≤ 20 mA), 40 V (20 mA < Io ≤ 50 mA), 20 V (50 mA < Io ≤ 100 mA), Io is the output current in A.