HPSMU Module Specifications  
Voltage range, resolution, and accuracy (high resolution ADC)  
Voltage  
range  
Force  
resolution  
Measure  
resolution  
Force accuracy 1  
±(% + mV)  
Measure accuracy 1  
±(% + mV)  
Maximum  
current  
±2 V  
100 µV  
1 mV  
2 mV  
5 mV  
10 mV  
2 µV  
±(0.018 + 0.4)  
±(0.018 + 3)  
±(0.018 + 6)  
±(0.018 + 15)  
±(0.018 + 30)  
±(0.01 + 0.14)  
±(0.009 + 0.9)  
±(0.01 + 1)  
±(0.012 + 2.5)  
±(0.014 + 2.8)  
1 A  
1 A  
500 mA  
125 mA  
50 mA  
±20 V  
±40 V  
±100 V  
±200 V  
20 µV  
40 µV  
100 µV  
200 µV  
1. ± (% of reading value + offset value in mV)  
Current range, resolution, and accuracy (high resolution ADC)  
Current  
range  
Force  
resolution  
Measure  
resolution  
Force accuracy 1  
±(% + A + A)  
Measure accuracy 1  
±(% + A + A)  
Maximum  
voltage  
±1 nA  
50 fA  
500 fA  
5 pA  
50 pA  
500 pA  
5 nA  
50 nA  
500 nA  
5 µA  
10 fA  
10 fA  
±(0.1 + 3E-13 + Vo x 1E-15)  
±(0.1 + 3E-12 + Vo x 1E-14)  
±(0.05 + 3E-11 + Vo x 1E-13)  
±(0.05 + 3E-10 + Vo x 1E-12)  
±(0.05 + 3E-9 + Vo x 1E-11)  
±(0.035 + 15E-9 + Vo x 1E-10)  
±(0.04 + 15E-8 + Vo x 1E-9)  
±(0.04 + 15E-7 + Vo x 1E-8)  
±(0.045 + 15E-6 + Vo x 1E-7)  
±(0.4 + 3E-4 + Vo x 1E-6)  
±(0.1 + 2E-13 + Vo x 1E-15)  
±(0.1 + 1E-12 + Vo x 1E-14)  
±(0.05 + 2E-11 + Vo x 1E-13)  
±(0.05 + 1E-10 + Vo x 1E-12)  
±(0.04 + 2E-9 + Vo x 1E-11)  
±(0.03 + 3E-9 + Vo x 1E-10)  
±(0.03 + 6E-8 + Vo x 1E-9)  
±(0.03 + 2E-7 + Vo x 1E-8)  
±(0.04 + 6E-6 + Vo x 1E-7)  
±(0.4 + 15E-5 + Vo x 1E-6)  
200 V  
200 V  
200 V  
200 V  
200 V  
200 V  
200 V  
200 V  
200 V2  
200 V2  
±10 nA  
±100 nA  
±1 µA  
±10 µA  
±100 µA  
±1 mA  
±10 mA  
±100 mA  
±1 A  
100 fA  
1 pA  
10 pA  
100 pA  
1 nA  
10 nA  
100 nA  
1 µA  
50 µA  
1. ± (% of reading value + fixed offset in A + proportional offset in A), Vo is the output voltage in V.)  
2. 200 V (Io ≤ 50 mA), 100 V (50 mA < Io ≤ 125 mA), 40 V (125 mA < Io ≤ 500 mA), 20 V (500 mA < Io ≤ 1 A), Io is the output current in A.  
Voltage range, resolution, and accuracy (high speed ADC)  
Voltage  
range  
Force  
resolution  
Measure  
resolution  
Force accuracy 1  
±(% + mV)  
Measure accuracy 1  
±(% + mV)  
Maximum  
current  
±2 V  
100 µV  
1 mV  
2 mV  
5 mV  
10 mV  
100 µV  
1 mV  
2 mV  
5 mV  
10 mV  
±(0.018 + 0.4)  
±(0.018 + 3)  
±(0.018 + 6)  
±(0.018 + 15)  
±(0.018 + 30)  
±(0.01 + 0.7)  
±(0.01 + 4)  
±(0.015 + 8)  
±(0.02 + 20)  
±(0.035 + 40)  
1 A  
1 A  
500 mA  
125 mA  
50 mA  
±20 V  
±40 V  
±100 V  
±200 V  
1. ±(% of reading value + offset value in mV). Averaging is 128 samples in 1 PLC.  
Current range, resolution, and accuracy (high speed ADC)  
Current  
range  
Force  
resolution  
Measure  
resolution  
Force accuracy 1  
±(% + A + A)  
Measure accuracy 1  
±(% + A + A)  
Maximum  
voltage  
±1 nA  
50 fA  
500 fA  
5 pA  
50 pA  
500 pA  
5 nA  
50 nA  
500 nA  
5 µA  
50 fA  
500 fA  
5 pA  
50 pA  
500 pA  
5 nA  
50 nA  
500 nA  
5 µA  
±(0.1 + 3E-13 + Vo x 1E-15)  
±(0.1 + 3E-12 + Vo x 1E-14)  
±(0.05 + 3E-11 + Vo x 1E-13)  
±(0.05 + 3E-10 + Vo x 1E-12)  
±(0.05 + 3E-9 + Vo x 1E-11)  
±(0.035 + 15E-9 + Vo x 1E-10)  
±(0.04 + 15E-8 + Vo x 1E-9)  
±(0.04 + 15E-7 + Vo x 1E-8)  
±(0.045 + 15E-6 + Vo x 1E-7)  
±(0.4 + 3E-4 + Vo x 1E-6)  
±(0.25 + 3E-13 + Vo x 1E-15)  
±(0.25 + 2E-12 + Vo x 1E-14)  
±(0.1 + 2E-11 + Vo x 1E-13)  
±(0.1 + 2E-10 + Vo x 1E-12)  
±(0.05 + 2E-9 + Vo x 1E-11)  
±(0.05 + 2E-8 + Vo x 1E-10)  
±(0.04 + 2E-7 + Vo x 1E-9)  
±(0.04 + 2E-6 + Vo x 1E-8)  
±(0.1 + 2E-5 + Vo x 1E-7)  
±(0.5 + 3E-4 + Vo x 1E-6)  
200 V  
200 V  
200 V  
200 V  
200 V  
200 V  
200 V  
200 V  
200 V2  
200 V2  
±10 nA  
±100 nA  
±1 µA  
±10 µA  
±100 µA  
±1 mA  
±10 mA  
±100 mA  
±1 A  
50 µA  
50 µA  
1. ±(% of reading value + fixed offset in A + proportional offset in A), Vo is the output voltage in V.)  
2. 200 V (Io ≤ 50 mA), 100 V (50 mA < Io ≤ 125 mA), 40 V (125 mA < Io ≤ 500 mA), 20 V (500 mA < Io ≤ 1 A), Io is the output current in A.  