09 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet
HCSMU Module Specifications
Voltage range, resolution, and accuracy
Voltage
range
Force
resolution
Measure
resolution
Force accuracy 1
±(% + mV + mV)
Measure accuracy 1
(% + mV + mV)
Maximum
current
±0.2 V
±2 V
±20 V
±40 V
200 nV
2 µV
20 µV
40 µV
200 nV
2 µV
20 µV
40 µV
±(0.06 + 0.14 + Io x 0.05)
±(0.06 + 0.6 + Io x 0.5)
±(0.06 + 3 + Io x 5)
±(0.06 + 0.14 + Io x 0.05)
±(0.06 + 0.6 + Io x 0.5)
±(0.06 + 3 + Io x 5)
20 A
20 A
20 A
1 A
±(0.06 + 3 + Io x 10)
±(0.06 + 3 + Io x 10)
1. ±(% of reading value + fixed offset in mV + proportional offset in mV). Note: Io is the output current in A.
Current range, resolution, and accuracy
Current
range
Force
resolution
Measure
resolution
Force accuracy 1
(% + A + A)
Measure accuracy 1
(% + A + A)
Maximum
voltage
±10 µA
±100 µA
±1 mA
10 pA
100 pA
1 nA
10 nA
100 nA
1 µA
10 pA
100 pA
1 nA
10 nA
100 nA
1 µA
±(0.06 + 2E-9 + Vo x 1E-10)
±(0.06 + 2E-8 + Vo x 1E-9)
±(0.06 + 2E-7 + Vo x 1E-8)
±(0.06 + 2E-6 + Vo x 1E-7)
±(0.06 + 2E-5 + Vo x 1E-6)
±(0.4 + 2E-4 + Vo x 1E-5)
±(0.4 + 2E-3 + Vo x 1E-4)
±(0.06 + 2E-9 + Vo x 1E-10)
±(0.06 + 2E-8 + Vo x 1E-9)
±(0.06 + 2E-7 + Vo x 1E-8)
±(0.06 + 2E-6 + Vo x 1E-7)
±(0.06 + 2E-5 + Vo x 1E-6)
±(0.4 + 2E-4 + Vo x 1E-5)
±(0.4 + 2E-3 + Vo x 1E-4)
40 V
40 V
40 V
40 V
40 V
40 V
20 V
±10 mA
±100 mA
±1 A
±20 A2
20 µA
20 µA
1. ±(% of reading value + fixed offset in A + proportional offset in A), Vo is the output voltage in V.
2. Pulse mode only. The maximum value of the base current during pulsing is ±100 mA.
Power consumption
HCSMU measurement and output range
Pulse only
Pulse only
Voltage source mode:
Pulse only
Current (A)
Voltage range
Power
DC and pulse
DC and pulse
DC and pulse
20
0.2 V
2 V
40 V
40 x Ic (W)
40 x Ic (W)
40 x Ic (W)
Where Ic is the current compliance setting.
For pulse current, Ic = (duty) x Ipulse
1
Current source mode:
Voltage
Power
compliance
-20
20
40
Voltage (V)
-40
Vc ≤ 0.2
0.2 < Vc ≤ 2
2 < Vc ≤ 40
40 x Io (W)
40 x Io (W)
40 x Io (W)
-1
Where Vc is the voltage compliance setting and
Io is output current.
For pulse current, Io = (duty) x Ipulse
-20
Current range expansion
If two HCSMUs are combined using the
Dual HCSMU combination adapter or the
Dual HCSMU Kelvin combination adapter,
then the maximum current ranges are 40
A (Pulsed) and 2 A (DC).