04 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet
B1505A Specifications
Supported plug-In modules
The B1505A supports ten slots for plug-in modules.
Part
Description
Slots
Range of operation
Measure
number
occupied
resolution
B1510A High Power Source Monitor Unit (HPSMU)
B1511B Medium Power Source Monitor Unit (MPSMU)
2
1
2
-200 V to 200 V, -1 A to 1 A
-100 V to 100 V, -100 mA to 100 mA
-40 V to 40 V, -1 A to 1 A
-20 V to 20 V, -20 A to 20 A (Pulse only)
-3000 V to 3000 V, -4 mA to 4 mA
-1500 V to 1500 V, -8 mA to 8 mA
-30 V to 30 V, -100 mA to 100 mA
-30 V to 30 V, -1 A to 1 A (Pulse only)
1 kHz to 5 MHz
2 µV, 10 fA
0.5 μV, 10 fA
200 nV, 10 pA
B1512A High Current Source Monitor Unit (HCSMU)
B1513C High Voltage Source Monitor Unit (HVSMU)
B1514A Medium Current Source Monitor Unit (MCSMU)
2
1
200 µV, 10 fA
200 nV, 10 pA
B1520A1 Multi Frequency Capacitance Measurement Unit (MFCMU) 1
1. N1300A-100 SMU CMU Unify Unit (SCUU) is not supported for the B1505A.
Maximum sink current: ± 4.2 A
Trigger in/out synchronization pulses
before and after setting and measuring
DC voltage and current. Arbitrary trig-
ger events can be masked or activated
independently.
Maximum module
configuration
Output terminal/connection:
Triaxial connector, Kelvin
(remote sensing)
The total power consumption of
all modules cannot exceed 84 W.
Under this rule, the B1505A can
contain any combination of the
following SMUs:
– Up to 4 dual-slot HPSMUs1
– Up to 10 single-slot MPSMUs
– Up to 2 dual-slot HCSMUs1
– Up to 6 single-slot MCSMUs
– Up to 5 dual-slot HVSMU
GNDU supplemental
characteristics
Furnished software
– EasyEXPERT group+
– VXI plug & play driver for the B1500A
– MDM file converter
This tool can convert the EasyEXPERT
group+ file (XTR/ZTR) to Keysight IC-CAP
MDM file format.
Load capacitance: 1 µF
Cable resistance:
For IS ≤ 1.6 A: Force line R < 1 Ω
For 1.6 A < IS ≤ 2.0 A: Force line R
< 0.7 Ω
For 2.0 A < IS ≤ 4.2 A: Force line R
< 0.35 Ω
For all cases: Sense line R ≤ 10 Ω
In addition, up to 1 single-slot MFCMU can
be installed per B1505A mainframe for any
of the above listed SMU configurations.
The EasyEXPERT file of the following mea-
surements performed in the classic mode
is only supported:
Where IS is the current being sunk
by the GNDU.
– IV Sweep
– Multi channel IV Sweep
– CV Sweep
The installation order of the modules
is: HPSMU, MPSMU, MFCMU, MCSMU,
HCSMU and HVSMU starting from the
bottom of the B1505A mainframe.
Peripherals and interface
Data storage
Hard disk drive, DVD-R drive
– 4155/56 setup file converter tool
This tool can convert 4155 and 4156 mea-
surement setup files (file extensions MES
or DAT) into equivalent EXPERT group+
classic test mode setup files
Maximum voltage between
Interfaces
GPIB, interlock, USB (USB 2.0, front 2,
rear 2), LAN (1000BASE-T/100BASE-
TX/10BASE-T), trigger in/out, digital I/O,
VGA video output
Common and Ground
≤ ± 42 V
Supported operating systems:
Ground unit (GNDU)
specifications
Microsoft Windows XP Professional
(Service Pack 3 or later), Windows Vista
Business (Service Pack 2 or later (32bit
only)), and Windows 7 Professional (Service
Pack 1 or later (32bit and 64bit)) Supported
language: English (US)) Supported .NET
Framework Microsoft .NET Framework 3.5
SP1
Remote control capabilities
– FLEX commands (GPIB)
– EasyEXPERT group+ remote control
function (LAN)
Trigger I/O
Only available using GPIB FLEX
commands.
The GNDU is furnished with the B1505A
mainframe.
Output voltage: 0 V ± 100 µV
1. The total number of installed HPSMU