Hardware  
Specification Condition  
The measurement and output accuracy are 3. Ambient temperature change less than  
The “supplemental” information and  
“typical” entries in the following speci-  
fications are not warranted, but provide  
useful information about the functions and  
performance of the instruments.  
specified at the rear panel connector termi-  
nals when referenced to the Zero Check  
terminal under the following conditions:  
1. 23 °C 5 °C (double between 5 °C to  
18 °C, and 28 °C to 40 °C if not noted  
otherwise)  
1 °C after auto calibration execution.  
4. Integration time: medium or long  
5. Filter: ON (for SMUs)  
6. Kelvin connection (for HRSMU, HPSMU,  
and GNDU)  
7. Calibration period: 1 year  
2. After 40 minutes warm-up  
Agilent 4156C Precision Semiconductor Parameter Analyzer  
Output Terminal/Connection:  
HRSMU (High Resolution SMU) Specifications  
Dual triaxial connectors, Kelvin (remote  
sensing)  
Voltage Range, Resolution, and Accuracy (HRSMU)  
Voltage Set Reso.  
Range  
Set Accuracy  
Meas.  
Reso.  
2 μV  
20 μV  
40 μV  
100 μV  
Meas. Accuracy  
Max.  
Current  
Voltage/Current Compliance  
(Limiting):  
2 V  
100 μV  
1 mV  
2 mV  
5 mV  
(0.02ꢀ%400 μV)  
(0.02ꢀ%3 mV)  
(0.025 ꢀ%6 mV)  
(0.03ꢀ%15 mV)  
(0.01ꢀ%200 μV)  
(0.01ꢀ%1 mV)  
(0.015ꢀ%2 mV)  
(0.02ꢀ%5 mV)  
100 mA  
20 V  
40 V  
100 V  
100 mA  
The SMU can limit output voltage or  
current to prevent damaging the device  
under test.  
1
2
1 100 mA (Vout ≤20 V), 50 mA (20 V<Vout≤40 V)  
2 100 mA (Vout ≤20 V), 50 mA (20 V<Vout≤40 V), 20 mA (40 V<Vout≤100 V)  
Voltage: 0 V to 100 V  
Current: 100 fA to 100 mA  
Compliance Accuracy: Same as the current  
(voltage) settling accuracy.  
Current Range, Resolution, and Accuracy (HRSMU)  
Current Set  
Set Accuracy  
Meas. Meas. Accuracy  
Reso.  
1 fA  
Max. V  
HRSMU Supplemental Information:  
Maximum allowable cable resistance when  
using Kelvin connection  
Range  
Reso.  
10 pA 10 fA  
100 pA 10 fA  
(4ꢀ%400 fA)1, 2  
(4ꢀ%20 fA%1 fAꢁVout/100)1, 2 100 V  
(4ꢀ%40 fA%10 fAꢁVout/100)1, 2 100 V  
(4ꢀ%400 fA)1, 2  
1 fA  
10 fA  
10 fA  
(Force, Sense): 10 Ω  
1 nA  
100 fA  
(0.5ꢀ%0.7 pA%1 fAꢁVout)2  
(0.5ꢀ%4 pA%10 fAꢁVout)  
(0.5ꢀ%0.4 pA%1 fAꢁVout)2  
(0.5ꢀ%2 pA%10 fAꢁVout)  
(0.1ꢀ%20 pA%100 fAꢁVout)  
(0.1ꢀ%200 pA%1 pAꢁVout)  
(0.05ꢀ%2 nA%10 pAꢁVout)  
100V  
100V  
100V  
100V  
100V  
Typical voltage source output resistance  
(Force line/non-Kelvin connection): 0.2 Ω  
Voltage measurement input resistance/  
current source output resistance:  
≥1015 Ω (10 pA range)  
10 nA 1 pA  
100 nA 10 pA  
(0.12ꢀ%40 pA%100 fAꢁVout) 100 fA  
1 μA  
100 pA (0.12ꢀ%400 pA%1 pAꢁVout) 1 pA  
10 μA 1 nA  
100 μA 10 nA  
(0.07ꢀ%4 nA%10 pAꢁVout)  
(0.07ꢀ%40 nA%100 pAꢁVout) 100 pA (0.05ꢀ%20 nA%100 pAꢁVout) 100V  
10 pA  
Current compliance setting accuracy for  
opposite polarity:  
1 mA  
10 mA 1 μA  
100 mA 10 μA  
100 nA (0.06ꢀ%400 nA%1 nAꢁVout) 1 nA  
(0.04ꢀ%200 nA%1 nAꢁVout)  
(0.04ꢀ%2 μA%10 nAꢁVout)  
100V  
100V  
(0.06ꢀ%4 μA%10 nAꢁVout)  
10 nA  
10 pA to 10 nA range: V/I setting  
accuracy 12ꢀ of range  
3
(0.12ꢀ%40 μA%100 nAꢁVout) 100 nA (0.1ꢀ%20 μA%100 nAꢁVout)  
100 nA to 100 mA range: V/I setting  
accuracy 2.5ꢀ of range  
1
2
The accuracy is applicable when offset cancellation has been performed.  
The offset current specification is multiplied by one of the following factors  
depending upon the ambient temperature and humidity (RH = Relative Humidity):  
Humidity ꢀ  
5 - 60  
ꢁ2  
RH  
60 - 80  
ꢁ2  
Temperature  
5 °C to 18 °C  
18 °C to 28 °C  
28 °C to 40 °C  
ꢁ1  
ꢁ2  
ꢁ2  
ꢁ5  
3
100 V (Iout≤20 mA), 40 V (20 mA<Iout≤50 mA), 20 V (50 mA<Iout≤100 mA)  
Vout is the output voltage in volts. Iout is the output current in amps.  
For example, accuracy specifications are given as ꢀ of set/measured  
value (0.04ꢀ) plus offset value (200 nA%1 nAꢁVout) for the 1 mA range.  
The offset value consists of a fixed part determined by the set/measure-  
ment range and a proportional part that is multiplied by Vout or Vout/100.  