12 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet
Pulse width resolution: 2 μs
Pulse period: 5 ms to 5 s
Pulse period resolution: 100 μs
Pulse duty:
SMU source
measurement mode
Power compliance
For HPSMU:
Power: 0.001 W to 20 W
Resolution: 0.001 W
For HPSMU and MPSMU:
VFIM, IFVM
For HCSMU, MCSMU
and HVSMU:
For 1 A range: ≤ 5%
For 10 μA to 100 mA range
Period ≥ delay + width + 2 ms
(when delay + width ≤ 100 ms)
Period ≥ delay + width + 10 ms
(when delay + width > 100 ms)
Pulse delay: 0 to (Period–width)
For MPSMU:
Power: 0.001 W to 2 W
Resolution: 0.001 W
VFIM, VFVM, IFVM, IFIM
For HCSMU:
Power: 0.001 W to 40 W (DC)
0.001 W to 400 W (Pulse)
Resolution: 0.001 W
Output terminal/connection:
For HVSMU:
For HPSMU and MPSMU:
Dual triaxial connector,
Kelvin (remote sensing)
Pulse width: 500 µs to 2 s
Pulse width resolution: 2 µs
Pulse period: 5 ms to 5 s
For MCSMU:
Power: 0.001 W to 3 W (DC)
0.001 W to 30 W (Pulse)
Resolution: 0.001 W”
For HCSMU:
Period ≥ delay + width + 2 ms (when
delay + width ≤ 100 ms)
Period ≥ delay + width + 10 ms (when
delay + width > 100 ms)
Triaxial connector (for sense) and coaxial
connector (for force)
For HVSMU:
No power compliance
Kelvin (remote sensing)
For MCSMU:
Pulse period resolution: 100 µs
Pulse delay: 0 to (Period – width)
Pulse output limitation:
SMU pulse measurement
Dual triaxial connector, Kelvin
(remote sensing)
Pulse width, period, and delay:
When the pulse voltage is more than
1500 V, the peak and base of pulse
should be same polarities.
Pulse measurement delay:
2 µs to (Period – pulse
For HVSMU:
For HPSMU and MPSMU:
Pulse width: 500 µs to 2 s
Pulse width resolution: 100 µs
Pulse period: 5 ms to 5 s
High voltage triaxial connector, non-
Kelvin
Period ≥ delay + width + 2 ms (when
delay + width ≤ 100 ms)
Period ≥ delay + width + 10 ms (when
delay + width > 100 ms)
Pulse period resolution: 100 µs
Pulse delay: 0 s
Voltage/current compliance
(limiting)
The SMU can limit output voltage or
current to prevent damaging the device
under test.
measurement time – 2 m) s,
2 µs resolution
Supplemental
Characteristics
Voltage:
For HCSMU:
0 V to ±200 V (HPSMU)
0 V to ±100 V (MPSMU)
0 V to ±40 V (HCSMU)
0 V to ±30 V (MCSMU)
0 V to ±3000 V (HVSMU)
Pulse width:
Current compliance setting ac-
curacy (for opposite polarity):
50 µs to 1 ms (20 A range)
50 µs to 2 s (10 µA to 1 A range)
Pulse width resolution: 2 µs
Pulse period: 5 ms to 5 s
Pulse period resolution: 100 µs
Pulse duty:
For HPSMU and MPSMU:
For 1 pA to 10 nA ranges:
V/I setting accuracy ±12% of range
For 100 nA to 1 A ranges:
V/I setting accuracy ±2.5% of range
For HCSMU and MCSMU:
For 10 µA to 1 A ranges:
V/I setting accuracy ±2.5% of range
For 20 A range (HCSMU):
V/I setting accuracy ±0.6% of range
For HVSMU:
Current:
±1 pA to ±1 A (HPSMU)
±1 pA to ±100 mA (MPSMU)
±10 nA to ±20 A (HCSMU)
±10 nA to ±1 A (MCSMU)
±1 pA to ±8 mA (HVSMU)
For 20 A range: ≤ 1%
For 10 µA to 1 A range
Period ≥ delay + width + 2 ms (when
delay + width ≤ 100 ms)
Period ≥ delay + width + 10 ms
(when delay + width > 100 ms)
Pulse delay: 0 to (Period–width)
Compliance accuracy:
Same as the current or voltage
set accuracy.
For MCSMU:
Pulse width:
For 1 nA to 10 nA ranges:
V/I setting accuracy ±12% of range
For 100 nA to 10 mA ranges:
V/I setting accuracy ±2.5% of range
10 μs to 100 ms (1 A range)
10 μs to 2 s (10 μA to 100 mA
range)