ES51967
6,600 Counts Auto DMM
mode. Each block is repeated twice in one conversion cycle. The detailed data format of
each packet is listed below.
4.1 FUNCTUON
This packet indicates the measurement mode of the meter. The following table
summarizes the transmitted code for each mode. Note that the encoding of this packet is
different from the encoding of FC1-FC5 switch.
Code
Measurement Mode VBAR=0, VASEL=0 VBAR=1,VASEL=0 VBAR=1,VASEL=1
0111011
30111101
30111111
0110000
0111001
0110011
Voltage
Auto µA Current
Auto mA Current
66 A current
Manual A Current
Ω
Don’t care
Auto µA Current
Auto mA Current
Auto 660.0A/6600A Auto 660.0V/6600V
Auto 66.00A/660.0A Auto 66.00V/660.0V
0110101
0110001
0110010
0110110
0110100
0111110
Continuity
Diode
Frequency
Capacitance
2Temperature
1ADP
Don’t care
1When the function code = 0110010, the measurement mode is determined by judge bit of Status packet.
If judge bit is 1, it means duty cycle mode. If judge bit is 0, it means ADP mode.
2When the function code = 0110100, the judge bit in the Status packet determines whether the unit is
Celsius or Fahrenheit.
3When the function code = 0111101 or 0111111, the measurement mode is determined by VBAR bit and
VASEL bit of OPTION1 packet.
4.2 RANGE
This packet indicates the full-scale range of the meter. When the meter operates in
continuity mode or diode mode, this packet is always 0110000 since the full-scale
ranges in these modes are fixed. The following table lists the code for each range in
each measurement mode.
Ω
Code
V
*2-range auto V/A
66 A Manual A ADP
Frequency Capacitor
0110000 6.600V Lower Range(IVSL) 66.00 A 6.600A ADP3 660.0Ω
660.0Hz
-
6.600nF
66.00nF
0110001 66.00V Higher Range(IVSH)
0110010 660.0V
0110011 1000V
0110100 660.0mV
0110101
66.00A ADP2 6.600KΩ
660.0A ADP1 66.00KΩ 6.600KHz 660.0nF
6600A ADP0 660.0KΩ 66.00KHz 6.600µF
6.600MΩ 660.0KHz 66.00µF
66.00MΩ 6.600MHz 660.0µF
66.00MHz 6.600mF
0110110
0110111
66.00mF
*It includes auto µA, mA, 66.00V/660.0V, 660.0V/1000V, 66.00A/660.0A, 660.0A/6000A.