ES51967
6,600 Counts Auto DMM
5.2 LCD display on condition
LCD Annunciator
Condition
V
A
Ω
F
In voltage measurement mode, and diode measurement mode.
In current measurement mode.
In resistance measurement mode, and continuity mode.
In capacitance measurement mode.
In continuity check mode.
In diode mode.
%
Hz
In duty cycle measurement.
In frequency mode.
ADP1
ADP2
ADP3
ADP4
DC
In ADP1 mode
In ADP2 mode
In ADP3 mode
In ADP4 mode
In DC voltage or DC current mode.
In AC voltage or AC current mode.
When automatic full scale range selection is enabled.
In manual mode.
When HOLD function is enabled.
When Relative function is enabled.
When Max or Min function is enabled.
When ZERO function is enabled.
In capacitor measurement mode and the full scale range is in the order of mF.
In capacitor measurement mode and the full scale range is in the order of uF.
AC
AUTO
MANU
HOLD
REL
Max / Min
ZERO
m1
µ1
n
m2
In capacitor measurement mode and the full scale range is in the order of nF.
In voltage or current measurement mode and the full scale range is in the order of 10-3.
In current measurement mode and the full scale range id in the order of uA.
µ2
M
K
In resistance measurement mode and the full scale range is in the order of MΩ
In resistance measurement mode and the full scale range is in the order of KΩ
In temperature measurement mode and when the unit is ℃
℃
℉
In temperature measurement mode and when the unit is ℉
When auto power off function is enabled.
When RS232 output is enabled.
APO
RS232
5.3 Operating Timing
ES51967 incorporates a dual slope ADC with four phases: ZI, AZ, INT and DINT.
The timing of each phase are listed below.
*Voltage / Ohm /ADP / Manual Current (for clampmeter) measurement:
Phase
ZI
Time
40ms
AZ
25ms
INT
DINT
100ms
165ms
P.S. In the DC voltage and ohm measurement with auto mode, if the range is
changed, the internal clock rate will increase four times and the new measurement
cycle becomes 1/4 times of the original cycle until the range is stable.