3196 Specifications
(Guaranteed accuracy period : 6 months / Certain specifications vary when measuring 400Hz circuits. Please inquire with your HIOKI distributor for details.)
Power factor
Measurement range
Measurement accuracy
-3. Input specifications
:
:
-1.000 (lead) to 0.000 to +1.000 (lag)
±1 dgt. from the calculation of each measurement value
(±3 dgt. for the sum)
Measurement line types
:
Single-phase 2-wire, Single-phase 3-wire, Three-phase 3-
wire (3P3W2M, 3P3W3M) or Three-phase 4-wire, plus one
extra input channel
Displacement power factor
Measurement method
Voltage :4 channels (U1 to U4) (channel U4 can be
switched between AC and DC)
Current :4 channels (I1 to I4)
Voltage between U1, U2, and U3 without inter-channel
isolation
Input channels
Input methods
:
:
:
Calculated from the phase difference between the
fundamental waveforms of voltage and current
-1.000 (lead) to 0.000 to +1.000 (lag)
Measurement range
Measurement accuracy
Voltage unbalance factor
Measurement method
:
:
±0.5% rdg. ±0.2% f.s. ±1 dgt.(±3 dgt. for the sum)
Voltage between U1 to U3 and U4 with inter-channel
isolation
:
:
Calculation for three-phase 3-wire (3P3W3M) and three
phase 4-wire fundamental waveforms of voltage
Current input by clamp-on sensor
Voltage : 4 MΩ ±10% (differential input)
Current : 200 kΩ ±10%
Simultaneous digital sampling of voltage and current
PLL synchronization
dropouts, so sampling is never interrupted)
Input resistance
:
:
Current unbalance factor
Measurement method
Calculation for three-phase 3-wire (3P3W3M) and
three-phase 4-wire fundamental waveforms of current
Measurement method
(automatically switches to fixed clock during
∆V10 flicker
Display items
:
:
∆V10, ∆V10 (average over one hour, fourth maximum over one hour,
maximum over one hour, overall maximum (during the measurement
Voltage at either U1, U2, or U3
42.5 to 69 Hz
PLL synchronization channel source
PLL synchronization frequency range
:
:
period)), ∆U (deviation with respect to nominal voltage)
Same operation as AGC for IEC flicker
Sampling frequency
:
Standard voltage: Auto
Measurement accuracy : ±2% rdg.
IEC flicker (short period flicker Pst, long period flicker Plt)
Measurement method
For calculations (including DC measurement)
For harmonic and inter-harmonic analysis
:
256 points/cycle
:256 points/8 cycles (for 400 Hz)
:
:
2048 points/10 cycles (for 50 Hz)
2048 points/12 cycles (for 60 Hz)
2048 points/80cycles (for 400 Hz)
2 MHz
: Per IEC61000-4-15
Pst is measured for 10 minutes, and Plt is measured for 2 hours
Measurement accuracy : ±5% rdg. or less of the limit value
Harmonic voltage, current and power (including fundamental waveform components)
For transient overvoltage (impulse)
A/D converter resolution :
For calculations (including DC measurement)
For transient overvoltage (impulse)
Voltage measurement range :
Analysis window
: Rectangular
: 1 to 50
:
:
:
16 bits
12 bits
Analysis orders
Measurement accuracy
Voltage/current
:
1st to 20th orders : ±0.5% rdg. ±0.2% f.s.
21st to 50th orders : ±1.0% rdg. ±0.3% f.s.
1st to 20th orders : ±0.5% rdg. ±0.2% f.s.
21st to 30th orders : ±1.0% rdg. ±0.3% f.s.
31st to 40th orders : ±2.0% rdg. ±0.3% f.s.
41st to 50th orders : ±3.0% rdg. ±0.3% f.s.
Channels 1 to 3
Channel 4
:150.00, 300.00, 600.00 Vrms
:60.000, 150.00, 300.00, 600.00 Vrms
±60.000, 600.00 V pk (DC measurement)
Power
:
3 or less
Current measurement range :
Voltage crest factor :
(for 50/60 Hz, clamp-on sensor accuracy must be included for current and power)
With Model 9694 sensor
With Model 9660 sensor
With Model 9661 sensor
With Model 9667 sensor
With Model 9669 sensor
:
5.0000, 50.000 Arms
50.000, 100.00 Arms
50.000, 500.00 Arms
50.000, 500.00 A or 500.00 A, 5.0000 kArms
100.00 A, 1.0000 kArms
:
:
:
:
Inter-harmonic voltage and current
Analysis window
Analysis orders
Harmonic voltage/current phase difference (including fundamental waveform content)
Measurement method
: Rectangular
: 0.5 to 49.5
4 or less
Current crest factor :
:
Difference between voltage and current phase angle
components
Display items
Measurement accuracy
:
:
Sum of all or multiple channels
1st to 3rd orders : ±2˚
4th to 50th orders : ±(0.02˚ × k+2˚), k = harmonic order
-4. Measurement specifications
(For specifications when measuring 400Hz circuits, please inquire with your HIOKI distributor.)
RMS voltage
(for 50/60 Hz, clamp-on sensor accuracy must be included for current and power)
Measurement method
:
True RMS (calculated continuously every 10 or 12 cycles at 50 or
60 Hz respectively)
-5. Display specifications
Range selection
:
:
Manual (channels 1 to 3 are set in the same operation)
AC : ±0.2% rdg. ±0.1% f.s.
DC : ±0.3% rdg. ±0.4% f.s.
Display device
:6.4" TFT color LCD (640 × 480 dots)
Measurement accuracy
Text display
:
English, German, French, Italian, Spanish, Chinese or
Japanese
RMS current
Measurement method
:
True RMS (calculated continuously every 10 or 12 cycles at 50 or
60 Hz respectively)
-6. External interface specifications
(1) External control terminals : External event input and output
Range selection
Measurement accuracy
:
:
Manual (channels 1 to 3 are set in the same operation)
±0.2% rdg. ±0.1% f.s. + clamp-on sensor accuracy
(2) PC card interface Slot
:
Compliant with PCMCIA/JEIDA PC Card Standard,
Type II slot × 1
Transient overvoltage (impulse)
Measurement method
Measurement range
Display items
:
:
:
:
2 MHz sampling
Compatible cards
(3) RS-232C interface
Standard
: Flash ATA cards up to 528 MB
2000 Vpk
4 ms waveform (2 ms before and after center peak)
Period exceeding threshold (max. 4 ms)
: EIA RS-232C-compliant (with 9-pin D-sub connector)
: Printer or modem or GPS
Period
Destination device
Minimum detectable duration
Measurement accuracy
:
0.5 µs
Printer interval selections : OFF, 1, 5, 10, or 30 minutes, 1 or 2 hours
(4) LAN interface
: ±5.0% rdg. ±20 V (1000 V DC and 700 Vrms/100 kHz)
Voltage swell (rise in RMS voltage)
Communications protocol : Ethernet and TCP/IP (with 10BASE-T RJ-45 connector)
Measurement method
:
True RMS (a single cycle is calculated by overlapping each half cycle)
(The voltage between lines is measured for three phase 3-wire lines, and
phase voltage is measured for three phase 4-wire lines.)
Amplitude and duration of swell
-7. Environment & safety specifications
Operating environment
Display items
Measurement accuracy
:
:
: Indoors, up to a height of 2000 m (6562.2 ft)
Same as RMS voltage
Storage temperature & humidity : -20 to 50˚C, max. 80% rh (non-condensating)
Operating temperature and humidity : 0 to 40˚C, max. 80% rh (non-condensating)
Maximum measurement terminal voltage : Voltage terminals : 780 Vrms AC,
1103 V peak
True RMS (a single cycle is calculated by overlapping each half cycle)
Current terminals : 1.7 Vrms AC,
2.4 V peak
Maximum in-phase voltage : 600 Vrms AC (50/60 Hz, voltage input terminals)
Withstand voltage
: 5.55 kVrms AC/15 sec
(50/60 Hz, 1 mA current sensitivity)
Between voltage and clamp input terminals, between the voltage
input terminal and 3196 casing, and between voltage input
terminals (U1 to U3) and voltage input terminal (U4)
: IP30 (per EN60529)
: EMC :EN61326:1997+A1:1998+A2:2001
CLASS A, EN61000-3-2:2000 and
EN61000-3-3:1995+A1:2001
Enclosure protection
Standards conformance
Safety :EN61010-1:2001
Voltage input unit :Contamination Level 2,
Measurement Category III
(Anticipated transient overvoltage: 6000 V)