33 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet
Accessories
Note: The total power consumption of all
modules cannot exceed 50 W when using
test fixture under the condition that operat-
ing temperature is more than 35 °C.
N1258A module selector
Supplemental characteristics
Leakage current:
For HPSMU:
Specifications
10 pA at 200 V
For HCSMU:
100 pA at 10 V (High Force to Low
Force, High Sense to Low Sense)
For HVSMU:
Input terminals:
HPSMU force port1, 1 ea., (Triaxial)
HPSMU sense port1, 1 ea., (Triaxial)
HCSMU force port, 1 ea. (BNC)
HCSMU sense port, 1 ea. (Triaxial)
HVSMU port2, 1 ea. (HV triaxial)
GNDU port, 1 ea. (Triaxial)
Digital I/O port, 1 ea. (D-sub 25 pin)
AC power line connector, 1 ea.
1. Either HPSMU or MPSMU can be
connected to HPSMU port.
2. Either HVSMU or HVMCU can be
connected to HVSMU port.
Output terminal:
Supplemental characteristics
Leakage current:
For HPSMU (Force, Sense) port:
10 pA at 200 V (Force, Sense)
For HCSMU (High Force, High sense) port:
100 pA at 10 V
For HVSMU (Force) port:
10 pA at 1500 V (humidity range: 20%
to 70% RH)
10 pA at 1500 V (humidity range: 20%
to 70% RH)
20 pA at 3000 V (humidity range: 20%
to 50% RH)
N1259A test fixture
20 pA at 3000 V (humidity range: 20%
to 50% RH)
Specifications
Input terminals:
HPSMU port1, 2 ea.
High force (HV triaxial)
High sense (HV triaxial)
Low force (BNC)
N1259A-010 inline package
socket module (3 pin)
Force, sense (Triaxial)
HCSMU port, 2 ea.
Force (BNC), sense (Triaxial)
HVSMU port2, 1 ea. (HV triaxial)
GNDU port, 1 ea. (Triaxial)
AUX port, 2 ea. (BNC)
Interlock port, 1 ea.
Low sense (BNC)
External relay control output
(D-sub 15 pin)
Protection:
HPSMU, GNDU, HCSMU Low Force
Power indicator:
LED turns yellow when AC power is
applied and turns green the module
selector is ready to use.
Status indicator:
Green LED lights to indicate the present
connection path of module selector;
Open, HCSMU, HPSMU, or HVSMU.
Maximum voltage/current:
For HPSMU port:
Specifications
Number of terminal:
Sockets, 6 ea. (Ø4 mm jack (banana))
DUT interface:
Inline package socket (3-pin)
Maximum voltage for terminals:
3000 Vdc
1. Either HPSMU or MPSMU can be con-
nected
to HPSMU port.
2. Either HVSMU or HVMCU can be con-
nected
to HVSMU port.
Protection:
HPSMU, GNDU, HCSMU Low Force
terminal
High voltage indicator:
LED turns red when a SMU output volt-
age is over 42 V.
Maximum voltage/current:
For HPSMU port:
N1259A-011 universal
socket module
Specifications
Number of terminal:
Sockets, 8 ea. (Ø4 mm jack (banana))
Maximum voltage for terminals:
3000 Vdc
±200 V/1 A
For HCSMU port:
±40 V/2 A, ±20 V/30 A
(Pulse width 1 ms, duty 1%)
For HVSMU port:
Force: ±200 V/1 A
Sense: ±200 V
For HCSMU port:
±3000 V/4 mA,
High Force: ±40 V/2 A, ±20 V/40 A
(Pulse width 1 ms, duty 1%)
Low Force: ±40 V/2 A, ±20 V/40 A
(Pulse width 1 ms, duty 1%)
High Sense: ±40 V
±1500 V/2.5 A, ±2200 V/1.1 A
Low Sense: ±40 V
For HVSMU port:
Force: ±3000 V/4 mA,
±1500 V/2.5 A, ±2200 V/1.1 A