DS1921H/Z  
Mission Samples Counter Register Map  
ADDR  
021Ah  
021Bh  
021Ch  
b7  
b6  
b5  
b4  
Low Byte  
Center Byte  
High Byte  
b3  
b2  
b1  
b0  
DEVICE SAMPLES COUNTER  
The Device Samples Counter indicates how many temperature measurements have taken place since the  
device was assembled at the factory. The value is stored as an unsigned 24-bit integer number. The  
maximum number that can be represented in this format is 16777215, which is higher than the expected  
lifetime of the DS1921H/Z iButton. This counter cannot be reset under software control.  
Device Samples Counter Register Map  
ADDR  
021Dh  
021Eh  
021Fh  
b7  
b6  
b5  
b4  
Low Byte  
Center Byte  
High Byte  
b3  
b2  
b1  
b0  
TEMPERATURE LOGGING AND HISTOGRAM  
Once setup for a mission, the DS1921H/Z logs the temperature measurements simultaneously byte after  
byte in the datalog memory as well as in histogram form in the histogram memory. The datalog memory  
is able to store 2048 temperature values measured at equidistant time points. The first temperature value  
of a mission is written to address location 1000h of the datalog memory, the second value to address  
location 1001h and so on. Knowing the starting time point (Mission Time Stamp), the interval between  
temperature measurements, the Mission Samples Counter, and the rollover setting, one can reconstruct  
the time and date of each measurement stored in the datalog.  
There are two alternatives to the way the DS1921H/Z will behave after the 2048 bytes of datalog memory  
is filled with data. With rollover disabled (RO = 0), the device will fill the datalog memory with the first  
2048 mission samples. Additional mission samples are not logged in the datalog, but the histogram, and  
temperature alarm memory continue to update. With rollover enabled (RO = 1), the datalog will wrap  
around, and overwrite previous data starting at 1000h for the every 2049th mission sample. In this mode  
the device stores the last 2048 mission samples.  
For the temperature histogram, the DS1921H/Z provides 64 bins that begin at memory address 0800h.  
Each bin consists of a 16-bit, non-rolling-over binary counter that is incremented each time a temperature  
value acquired during a mission falls into the range of the bin. The least significant byte of each bin is  
stored at the lower address. Bin 0 begins at memory address 0800h, bin 1 at 0802h, and so on up to  
087Eh for bin 63, as shown in Figure 7. The number of the bin to be updated after a temperature  
conversion is determined by cutting off the two least significant bits of the binary temperature value. Out  
of range values are range locked and counted as 00h or FFh.  
12 of 44  