16 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Table 18. Test port input (dynamic accuracy)
Accuracy of the test port input power reading is relative to –10 dBm reference input power level.
Description
Specification
± 0.303 dB
(at 10 dBm)
Typical
Dynamic accuracy magnitude
(Option 3L3/3L4/3L5)
Reference = –10 dBm
± 0.087 dB
(at –30 dBm)
± 2.141 dB
(at –100 dBm)
± 0.303 dB
(at 10 dBm)
Dynamic accuracy magnitude
(Option 115, 135, 215, 235, 117, 137, 217, 237)
Reference = –10 dB
± 0.087 dB
(at –30 dBm)
± 2.141 dB
(at –100 dBm)
300 kHz to 3 GHz
± 0.383 dB
(at 10 dBm)
± 0.167 dB
(at –30 dBm)
± 2.221 dB
(at –100 dBm)
100 to 300 kHz
± 2.04 °
Dynamic accuracy phase
(Option 3L3/3L4/3L5)
Reference = –10 dB
(at 10 dBm)
± 0.58 °
(at -30 dBm)
± 16.23 °
(at –100 dBm)
± 2.04 °
Dynamic accuracy phase
(Option 115, 135, 215, 235, 117, 137, 217, 237)
Reference = –10 dB
(at 10 dBm)
± 0.58 °
(at –30 dBm)
± 16.23 °
(at –100 dBm)
300 kHz to 3 GHz
± 2.58 °
(at 10 dBm)
± 1.11 °
(at –30 dBm)
± 16.94 °
(at –100 dBm)
100 to 300 kHz