13 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet  
Test Port Output (Source)  
Table 11. Test port output frequency  
Description  
Specification  
5 Hz to 500 MHz  
5 Hz to 1.5 GHz  
5 Hz to 3 GHz  
100 kHz to 3 GHz  
100 kHz to 1.5 GHz  
1 mHz  
Typical  
Range (Option 3L3)  
Range (Option 3L4)  
Range (Option 3L5)  
Range (Option 115, 135, 215, 235, 137, 237)  
Range (Option 135, 235, 137, 237)  
Resolution  
Source stability  
± 7 ppm (5 to 40 °C)  
CW accuracy  
High stability option (Option 1E5)  
CW accuracy  
± 7 ppm ± 1 mHz  
± 1 ppm ± 1 mHz  
Stability  
± 0.05 ppm (5 to 40 °C)  
± 0.5 ppm per year  
Table 12. Test port output power  
Description  
Specification  
Typical  
Level accuracy  
(Option 3L3/3L4/3L5)  
± 0.8 dB (at 0 dBm, 50 MHz absolute)  
± 1.0 dB (at 5 Hz to 1.5 GHz, 0 dBm,  
relative to 50 MHz)  
± 1.5 dB (at 1.5 GHz to 3 GHz, 0 dBm,  
relative to 50 MHz)  
Level accuracy  
(Option 115, 135, 215,  
235)  
± 0.8 dB (at 0 dBm, 50 MHz absolute)  
± 1.5 dB (at 100 kHz to 300 kHz, 0 dBm,  
relative to 50 MHz)  
± 1.0 dB (at 300 kHz to 3 GHz, 0 dBm,  
relative to 50 MHz)  
Level accuracy  
(Option 117, 137, 217,  
237)  
± 0.8 dB (at 0 dBm, 50 MHz absolute)  
± 1.5 dB (at 100 kHz to 300 kHz, 0 dBm,  
relative to 50 MHz)  
± 1.0 dB (at 2 GHz to  
3 GHz, 0 dBm, relative  
to 50 MHz)  
± 1.0 dB (at 300 kHz to 2 GHz, 0 dBm,  
relative to 50 MHz)  
Level lineality  
± 0.75 dB (at –10 to 10 dBm, 0 dBm reference)  
(Option 3L3/3L4/3L5)  
Level lineality  
(Option 115, 135, 215,  
± 0.75 dB (at –10 to 5 dBm, 100 kHz to 300 kHz,  
0 dBm reference)  
235, 117, 137, 217, 237) ± 0.75 dB (at –10 to 10 dBm, 300 kHz to 3 GHz,  
0 dBm reference)  
Range  
–45 dBm to 10 dBm  
(Option 3L3/3L4/3L5)  
Range  
–45 dBm to 5 dBm (at 100 kHz to 300 kHz)  
–45 dBm to 10 dBm (at 300 kHz to 3 GHz)  
(Option 115, 135, 215,  
235, 117, 137, 217, 237)  
Sweep range  
–45 dBm to 10 dBm  
(Option 3L3/3L4/3L5)  
Sweep range  
–45 dBm to 5 dBm (at 100 kHz to 300 kHz)  
–45 dBm to 10 dBm (at 300 kHz to 3 GHz)  
(Option 115, 135, 215,  
235, 117, 137, 217, 237)  
Level resolution  
0.05 dB  
Table 13. Test port output signal purity  
Description  
Specification  
Typical  
Harmonics (2nd or 3rd)  
< -20 dBc (at 100 kHz to 300 kHz, 5 dBm)  
< -25 dBc (at 300 KHz to 3 GHz, 5 dBm)  
< –25 dBc (at 5 dBm)  
Non-harmonic spurious  