15 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Table 16. Test port input (trace noise)
Description
Specification
Typical
Trace noise magnitude
5 mdB rms
(Option 3L3/3L4/3L5)
(< 10 kHz)
source power level = +10 dBm
Automatic IF bandwidth
5 mdB rms
(10 kHz to 3 GHz)
3 kHz bandwidth
Trace noise magnitude
(Option 115, 135, 215, 235, 117,
137, 217, 237)
15 mdB rms
(100 to 300 kHz)
Maximum output power level
3 kHz Bandwidth
8 mdB rms
(300 kHz to 1 MHz)
5 mdB rms
(1 MHz to 3 GHz)
Trace noise phase
0.03° rms
(Option 3L3/3L4/3L5)
source power level = +10 dBm
(< 10 kHz)
Automatic IF bandwidth
0.03° rms
(10 kHz to 3 GHz)
3 kHz Bandwidth
Trase noise phase
0.09° rms
(Option 115, 135, 215, 235, 117,
137, 217, 237)
(100 to 300 kHz)
Maximum output power level
3 kHz Bandwidth
0.05° rms
(300 kHz to 1 MHz)
0.03° rms
(1 MHz to 3 GHz)
Table 17. Test port input (stability)
Description
Specification
SPD
Stability magnitude
(Option 3L3/3L4/3L5)
3 MHz to 3 GHz
0.01 dB/°C
Stability magnitude (Option 115,
135, 215, 235, 117, 137, 217, 237)
100 kHz to 300 kHz
300 kHz to 3 MHz
3 MHz to 3 GHz
0.05 dB/°C
0.02 dB/°C
0.01 dB/°C
Stability phase
(Option 3L3/3L4/3L5)
3 MHz to 3 GHz
0.1°/°C
Stability phase (Option 115, 135,
215, 235, 117, 137, 217, 237)
100 kHz to 300 kHz
300 kHz to 3 MHz
3 MHz to 3 GHz
0.5°/°C
0.2°/°C
0.1°/°C