14 | Keysight | E5061B Network Analyzer, 100 kHz to 1.5/3 GHz, 5 Hz to 500 M/1.5 G/3 GHz - Data Sheet
Test Port Input
Table 14. Test port input levels
Description
Specification
Typical
Damage level
+20 dBm, ± 7 V DC
(Option 3L3/3L4/3L5)
Damage level (Option 115, 135,
215, 235, 117, 137, 217, 237)
+20 dBm, ± 30 V DC
Absolute amplitude accuracy
< ± 3 dB
(at 0 dBm)
Crosstalk 1
(Option 3L3/3L4/3L5)
–85 dB (at 5 Hz to 100 Hz)
–100 dB (at 100 Hz to 9 kHz)
–110 dB (at 9 to 100 kHz)
–115 dB (at 100 kHz to 3 GHz)
Crosstalk 2
(Option 115, 135, 215, 235,
117, 137, 217, 237)
–100 dB (at 100 to 300 kHz)
–110 dB (at 300 k to 1MHz)
–115 dB (at 1 MHz to 3 GHz)
Table 15. Test port input (noise floor)
Description
Specification
Typical
Noise floor
(Option 3L3/3L4/3L5)
3 kHz bandwidth
–80 dBm (100 k to 1 MHz)
–85 dBm (1M to 3 GHz)
Noise floor (Option 3L3/3L4/3L5)
10 Hz bandwidth
–80 dBm (5 to 100 Hz)
–90 dBm (100 Hz to 9 kHz)
–100 dBm (9 to 100 kHz)
–105 dBm (100 k to 1 MHz)
–110 dBm (1 M to 3 GHz)
Noise floor (Option 115, 135, 215,
235, 117, 137, 217, 237)
3 kHz bandwidth
–70 dBm (100 to 300 kHz)
–80 dBm (300 k to 1 MHz)
–85 dBm (1M to 3 GHz)
Noise floor (Option 115, 135, 215,
235, 117, 137, 217, 237)
10 Hz bandwidth
–95 dBm (100 to 300 kHz)
–105 dBm (300 k to 1 MHz)
–110 dBm (1M to 3 GHz)
1. The specification might not be met at the frequencies 25 MHz and 90 MHz Line and Fan related
frequency.
2. Maximum test port input level: +10 dBm