21 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Table 32. Test signal current  
Range  
0 Arms - 100 mArms  
Resolution  
1 µArms (0 Arms - 2 mArms)  
2 µArms (2 mArms - 5 mArms)  
5 µArms (5 mArms - 10 mArms)  
10 µArms (10 mArms - 20 mArms)  
20 µArms (20 mArms - 50 mArms)  
50 µArms (50 mArms - 100 mArms)  
Setup accuracy  
normal  
±(10% + 10 µArms) (test signal voltage ≤ 20 mArms)  
(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)  
±(10% + 100 µArms) (test frequency ≤ 300 kHz,  
test signal current > 20 mArms) (spec.)  
±(15% + 200 µArms) (test frequency > 300 kHz,  
test signal voltage > 20 mArms) (test frequency ≤ 1 MHz : spec.,  
test frequency > 1 MHz : typ.)  
Constant1  
±(6% + 10 µArms) (test signal voltage ≤ 20 mArms)  
(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)  
±(6% + 100 µArms) (test frequency ≤ 300 kHz,  
test signal voltage > 20 mArms) (spec.)  
±(12% + 200 µArms) (test frequency > 300 kHz,  
test signal voltage > 20 mArms) (test frequency ≤ 1 MHz : spec.,  
test frequency > 1 MHz : typ.)  
Test signal level monitor function  
– Test signal voltage and test signal current can be monitored.  
– Level monitor accuracy:  
Table 33. Test signal voltage monitor accuracy (Vac)  
Test signal voltage2  
Test frequency  
≤ 1 MHz  
Specification  
5 mVrms to 2 Vrms  
±(3% of reading value + 0.5 mVrms)  
±(6% of reading value + 1 mVrms)  
±(3% of reading value + 5 mVrms)  
±(6% of reading value + 10 mVrms)3  
> 1MHz  
> 2 Vrms  
≤ 300 kHz  
> 300 kHz  
Table 34. Test signal current monitor accuracy (Iac)  
Test signal current2  
Test frequency  
≤ 1 MHz  
Specification  
50 µArms to 20 mArms  
±(3% of reading value + 5 µArms)  
±(6% of reading value + 10 µArms)  
±(3% of reading value + 50 µArms)  
±(6% of reading value + 100 µArms)  
> 1MHz  
> 20 mArms  
≤ 300 kHz  
> 300 kHz  
1. When auto level control function is on.  
2. This is not an output value but a displayed test  
signal level  
3. Typ. when test frequency is > 1 MHz with test  
signal voltage > 10 Vrms.  