17 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Compensation function  
Table 28. The E4980A provides three types of compensation functions:  
OPEN compensation, SHORT compensation, and LOAD compensation.  
Type of compensation  
Description  
OPEN compensation  
Compensates errors caused by the stray admittance (C, G) of the test  
fixture.  
SHORT compensation  
LOAD compensation  
Compensates errors caused by the residual impedance (L, R) of the test  
fixture.  
Compensates errors between the actual measured value and a known  
standard value under the measurement conditions desired by the user.  
List sweep  
Points: There is a maximum of 201 points.  
First sweep parameter (primary parameter): Test frequency, test signal voltage, test  
signal current, test signal voltage of DC bias signal, test signal current of DC bias  
signal, DC source voltage.  
Note  
A parameter selected for one of the two  
parameters cannot be selected for the other  
parameter. It is not possible to set up a  
combination of test signal voltage and test  
signal current or one of test signal voltage of DC  
bias signal and test signal current of DC bias.  
Second sweep parameter (secondary parameter): None, impedance range, test  
frequency, test signal voltage, test signal current, test signal voltage of DC bias signal,  
test signal current of DC bias signal, DC source voltage  
The secondary parameter can be set only with  
SCPI commands.  
Trigger mode  
Sequential mode: When the E4980A is triggered once, the device is measured at all  
sweep points. /EOM/INDEX is output only once.  
Step mode: The sweep point is incremented each time the E4980A is triggered.  
/EOM/INDEX is output at each point, but the result of the comparator function of the  
list sweep is available only after the last /EOM is output.  