欢迎访问ic37.com |
会员登录 免费注册
发布采购

E4980A-002 参数 Datasheet PDF下载

E4980A-002图片预览
型号: E4980A-002
PDF下载: 下载PDF文件 查看货源
内容描述: [Precision LCR Meter]
分类和应用:
文件页数/大小: 38 页 / 1698 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号E4980A-002的Datasheet PDF文件第14页浏览型号E4980A-002的Datasheet PDF文件第15页浏览型号E4980A-002的Datasheet PDF文件第16页浏览型号E4980A-002的Datasheet PDF文件第17页浏览型号E4980A-002的Datasheet PDF文件第19页浏览型号E4980A-002的Datasheet PDF文件第20页浏览型号E4980A-002的Datasheet PDF文件第21页浏览型号E4980A-002的Datasheet PDF文件第22页  
18 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Comparator function of list sweep: The comparator function enables setting one pair of  
lower and upper limits for each measurement point.  
You can select from: Judge with the first sweep parameter/Judge with the second  
parameter/Not used for each pair of limits.  
Time stamp function: In the sequential mode, it is possible to record the measurement  
starting time at each measurement point by defining the time when FW detects a trigger  
as 0 and obtain it later with the SCPI command.  
Comparator function  
Bin sort: The primary parameter can be sorted into 9 BINs, OUT_OF_BINS, AUX_BIN,  
and LOW_C_REJECT. The secondary parameter can be sorted into HIGH, IN, and LOW.  
The sequential mode and tolerance mode can be selected as the sorting mode.  
Limit setup: Absolute value, deviation value, and % deviation value can be used for setup.  
BIN count: Countable from 0 to 999999.  
DC bias signal  
Table 29. Test signal voltage  
Range  
0 V to +2 V  
Resolution  
Accuracy  
0 V / 1.5 V / 2 V only  
0.1% + 2 mV (23 °C ± 5 °C)  
(0.1% + 2 mV) × 4  
(0 to 18 °C or 28 to 55 °C)  
Output impedance:  
100 Ω (nominal)  
Measurement assistance functions  
Note  
Data buffer function: Up to 201 measurement results can be read out in a batch.  
The following USB memory can be used.  
Complies with USB 1.1; mass storage class,  
FAT16/FAT32 format; maximum consumption  
current is below 500 mA.  
Save/Recall function:  
– Up to 10 setup conditions can be written to/read from the built-in non-volatile  
memory.  
– Up to 10 setup conditions can be written to/read from the USB memory.  
– Auto recall function can be performed when the setting conditions are written to  
Register 10 of the USB memory.  
Recommended USB memory: 4 GB USB Flash  
memory (Keysight PN 1819-0637).  
Use the recommended USB memory device  
exclusively for the E4980A, otherwise,  
previously saved data may be cleared.  
If you use a USB memory other than the  
recommended device, data may not be  
saved or recalled normally.  
Key lock function: The front panel keys can be locked.  
GPIB: 24-pin D-Sub (Type D-24), female; complies with IEEE488.1, 2 and SCPI  
Keysight will NOT be responsible for data loss in  
the USB memory caused by using the E4980A.  
USB host port: Universal serial bus jack, type-A (4 contact positions, contact 1 is on  
your left), female (for connection to USB memory only).  
USB interface port: Universal serial bus jack, type mini-B (4 contact positions); complies  
with USBTMC-USB488 and USB 2.0; female; for connection to the external controller.  
USBTMC: Abbreviation for USB Test & Measurement Class  
LAN: 10/100 BaseT Ethernet, 8 pins (two speed options)  
LXI Compliance: Class C (only applies to units with firmware revision A.02.00 or later)  
 复制成功!