20 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet  
Power and DC bias enhancement specification  
Increases test signal voltage and adds the variable DC bias voltage function.  
The Vdc-Idc measurement function is available when the option 001 is installed.  
Measurement parameters  
The following parameters can be used.  
– Lp-Rdc  
– Ls-Rdc  
– Vdc-Idc  
where  
Rdc  
Vdc  
Idc  
Direct-current resistance (DCR)  
Direct-current voltage  
Direct-current electricity  
Test signal  
Signal level  
Table 31. Test signal voltage  
Range  
0 Vrms to 20 Vrms (test frequency ≤ 1 MHz)  
0 Vrms to 15 Vrms (test frequency > 1 MHz)  
Resolution  
100 µVrms (0 Vrms - 0.2 Vrms)  
200 µVrms (0.2 Vrms - 0.5 Vrms)  
500 µVrms (0.5 Vrms - 1 Vrms)  
1 mVrms (1 Vrms - 2 Vrms)  
2 mVrms (2 Vrms - 5 Vrms)  
5 mVrms (5 Vrms - 10 Vrms)  
10 mVrms (10 Vrms - 20 Vrms)  
Setup accuracy  
normal ± (10% + 1 mVrms) (test signal voltage ≤ 2 Vrms)  
(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)  
± (10% + 10 mVrms) (Test frequency ≤ 300 kHz,  
test signal voltage > 2 Vrms) (spec.)  
± (15% + 20 mVrms) (test frequency > 300 kHz,  
test signal voltage > 2 Vrms) (test frequency ≤ 1 MHz : spec., test  
frequency > 1 MHz : typ.)  
Constant1 ± (6% + 1 mVrms) (test signal voltage ≤ 2 Vrms)  
(test frequency ≤ 1 MHz : spec., test frequency > 1 MHz : typ.)  
± (6% + 10 mVrms) (test frequency ≤ 300 kHz,  
test signal voltage > 2 Vrms) (spec.)  
± (12% + 20 mVrms) (test frequency > 300 kHz,  
test signal voltage > 2 Vrms) (test frequency ≤ 1 MHz : spec.,  
test frequency > 1 MHz : typ.)  
1. When auto level control function is on.  