37 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
Search, navigate, and lister
Type
Edge, pulse width, rise/fall, runt, frequency peak, serial bus 1, serial bus 2
Copy
Copy to trigger, copy from trigger
Frequency peak
Source
Math functions
Max number of peaks
Control
11
Threshold, excursion, results order (frequency or amplitude)
Event lister or navigation. Manual or autoscroll via navigation or touch event lister entry to jump to a
specific event
Result display
Waveform measurements
DC vertical accuracy/cursors 2
Single cursor accuracy: ± [DC vertical gain accuracy + DC vertical offset accuracy + 0.21% full scale]
Dual cursor accuracy: ± [DC vertical gain accuracy + 0.42% full scale] 1
56 automatic measurements, maximum of 10 displayed at a time
2 pairs of XY cursors
Number of measurements
Cursors
Automatic measurement of positions, ΔX, 1/ΔX, ΔY, and ΔY/ΔX
Measurements continuously updated with statistics. Cursors track last selected measurement.
Select up to 10 measurements from the list below:
Automatic
measurements
Snapshot
Makes a snapshot of 31 most popular measurements. Touchable target to populate the
measurement side bar
Voltage/current
Time
Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, preshoot, average- N cycles,
average-full screen, DC RMS-N cycles, DC RMS-full screen, AC RMS-N cycles, AC RMS-full screen
(standard deviation), ratio-N cycles, ratio-full screen
Period, frequency, counter, + width, - width, burst width, + duty cycle, - duty cycle, bit rate, rise
time, fall time, delay, phase, X at min Y, X at max Y
Count
Mixed
Jitter
Positive pulse count, negative pulse count, rising edge count, falling edge count
Area-N cycles, area-full screen
Option DSOX6JITTER required
Data TIE, clock TIE, N-period, period-period, + width to + width, - width to - width
Option DSOX6JITTER required
Real-time eye
Eye width, eye height
Dual-channel (requires Charge-N cycles, charge-full screen, peak-peak, amplitude, DC RMS-N cycles, DC RMS-full screen,
N2820A probe)
AC RMS- N cycles, AC RMS-full screen (standard deviation), average-N cycles, average-full screen,
base
Automatic
Available via BV0004B BenchVue
measurement logging
Counter
Built-in frequency counter (see “Precision counter/totalizer section” for the 10-digit counter)
Source
Resolution
Max frequency
Any analog and digital channel
5 digits
1 GHz (1.2 GHz typical)
1. Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
2. 1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.