 浏览型号DSOX6B10T252BW的Datasheet PDF文件第31页浏览型号DSOX6B10T252BW的Datasheet PDF文件第32页浏览型号DSOX6B10T252BW的Datasheet PDF文件第33页浏览型号DSOX6B10T252BW的Datasheet PDF文件第34页浏览型号DSOX6B10T252BW的Datasheet PDF文件第36页浏览型号DSOX6B10T252BW的Datasheet PDF文件第37页浏览型号DSOX6B10T252BW的Datasheet PDF文件第38页浏览型号DSOX6B10T252BW的Datasheet PDF文件第39页  
35 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet  
Performance Characteristics (Continued)  
Trigger system  
Trigger jitter  
< 1.0-ps rms with the jitter-free trigger  
< 3.0-ps rms without the jitter-free trigger  
Trigger bandwidth  
Edge  
Other modes  
1 GHz bandwidth  
500 MHz, 1 GHz, 2.5 GHz models: bandwidth of oscilloscope. 4-GHz and 6-GHz models: 3.5 GHz  
Bandwidth of oscilloscope or 1 GHz, whichever is smaller  
Trigger sensitivity  
(internal) 1  
≤ 10 mV/div  
> 10 mV/div  
≤ 10 mV/div  
DC to 1 GHz  
DC to 1 GHz  
DC to 2 GHz  
2.0 to 3.5 GHz  
DC to 2 GHz  
2.0 to 3.5 GHz  
Greater of 1 div or 5 mVpp  
0.6 div  
Greater of 1 div or 5 mVpp  
Greater of 1.5 div or 5 mVpp  
0.6 div  
2.5, 4, and 6 GHz  
bandwidth  
> 10 mV/div  
1.0 div  
Trigger sensitivity  
(external) 1  
Trigger level range  
± 1.6 V  
± 8 V  
Any channel  
External  
40 mVpp DC to 100 MHz, 70 mVpp 100 to 200 MHz  
200 mVpp DC to 100 MHz, 350 mVpp 100 to 200 MHz  
± 6 div from center screen  
8-V range = ± 8 V; 1.6-V range = ± 1.6 V  
1. Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration  
temperature.  
Trigger type selections  
InfiniiScan Zone (hardware zone qualifier)  
Trigger on user-defined zones drawn on the display. Applies to one analog channel at a time.  
Specify zones as either “must intersect” or “must not intersect.” Up to two zones. > 160,000 wfm/  
sec update rate  
Supported modes: Normal, peak detect, high resolution, max update rate  
Also works simultaneously with the serial decodes and mask/limit test  
Trigger on a rising and falling edge of any source, alternating or either edge of analog and digital  
channels  
Edge  
Edge then edge (B trigger)  
Pulse width  
Arm on a selected edge, wait a specified time, then trigger on a specified count of another selected  
edge. Minimum 4 ns  
Trigger on a pulse on a selected channel, whose time duration is less than a value, greater than a  
value, or inside a time range  
Minimum duration setting: 2 ns  
Maximum duration setting: 10 s  
Range minimum: 10 ns  
Pattern  
Trigger when a specified pattern of high, low, and don’t-care levels on any combination of analog,  
digital, or trigger channels is [entered | exited]. Pattern must have stabilized for a minimum of 2 ns to  
qualify as a valid trigger condition  
Minimum duration setting: 2 ns  
Maximum duration setting: 10 s  
Or  
Trigger on any selected edges from available sources (analog and digital channels only).  
Bandwidth is 500 MHz  
 复制成功!