Amplitude
Range: Minimum
20 mVpp if |offset| ≤ 0.5 Vpp into Hi-Z 1
10 mVpp if |offset| ≤ 0.5 Vpp into 50 Ω 1
10 Vpp except, 9 Vpp if sinc or cardiac, 7.5 Vpp if Gaussian pulse into Hi-Z;
5 Vpp/4.5 Vpp into 50 Ω
Range: Maximum
Resolution
Accuracy
Range
100 μV or 3 digits, whichever is higher
1.5% (frequency = 1 kHz)
± 5 V into Hi-Z, except ± 4 V if sine wave, ± 2.5 V if sinc, cardiac, or Gaussian
pulse into Hi-Z
DC offset
± 2.5 V into Hi-Z, except ± 2 V if sine wave, ± 1.25 V if sinc, cardiac, or
Gaussian pulse into 50 Ω
Resolution
Larger of 250 uV or 3 digits
Accuracy (waveform modes)
Accuracy (DC mode)
Impedance
± 1.5% of offset setting ± 1% of amplitude ± 1 mV
± 1.5% of offset setting ± 3 mV
50 Ω typical
Main output
Isolation
Protection
Not available, main output BNC is grounded
Overload automatically disables output
Trigger output
Trigger output available on trig-out BNC
Quick action customization key
Quick measure all
Quick measure statistics reset
Quick mask statistics reset
Quick histogram statistics reset
Quick print
Displays a popup containing a snapshot of all the single waveform measurements
Resets all measurement statistics and the measurement count
Resets mask statistics and counters
Zeros the histogram counters
Print the current screen image
Quick save
Quick email
Quick recall
Saves the current setup, screen image, or data file as specified in the settings menu
Emails the current setup, screen image, or data file as specified in the settings menu
Recalls setup, mask, or reference waveform
Quick freeze display
Freezes the display without stopping running acquisitions or unfreezes the display if currently frozen.
Waveform intensity preserved
Quick trigger mode
Quick clear display
Toggles the trigger mode between auto and normal
Clears the display
Display characteristics
Display
Display mode
Resolution
Graticules
12.1-inch capacitive multi-touch/gesture enabled color TFT LCD
Zone/zoom/annotation mode and waveform placement mode
800 (H) x 600 (V) pixel format (screen area)
8 vertical divisions by 10 horizontal divisions with intensity controls
YT and XY
Format
Maximum waveform update rate
> 135,000 wfm/s (real time)
> 450,000 wfm/s (real time max update rate)
Off, infinite, variable persistence (100 ms to 60 s)
256 intensity levels
Persistence
Intensity gradation
1. Sinc, cardiac and Gaussian pulse: ± 1.25 V into Hi-Z; ± 625 mV into 50 Ω.