33 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
DSOX/MSO 6000 X-Series digital storage/mixed signal oscilloscopes (Continued)
Vertical system digital channels
Digital input channels
Thresholds
16 digital (D0 to D15. Pod 1: D7 ~ D0, Pod 2: D15 ~ D8)
Threshold per pod
Threshold selections
User-defined threshold range
Maximum input voltage
Threshold accuracy 1
Maximum input dynamic range
Minimum voltage swing
Input impedance
TTL (+1.4 V), 5V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod)
± 8.0 V in 10-mV steps
± 40 V peak
± (100 mV + 3% of threshold setting)
± 10 V about threshold
500 mVpp
100 kΩ ± 2% at probe tip
~8 pF
Input capacitance
Vertical resolution
1 bit
Horizontal system analog channels
1 GHz
2.5 GHz
4 GHz
6 GHz
Time base range
Time base accuracy 1
500 ps/div to 50 s/div
± 1.6 ppm + aging factor
100 ps/div to 50 s/div
100 ps/div to 50 s/div
100 ps/div to 50 s/div
(1 year: ± 0.5 ppm, 2 years: ± 0.7 ppm, 5 years: ± 1.5 ppm, 10 years: ± 2.0 ppm)
Time base resolution
2.5 ps
Time base delay time
range
Pre-trigger
Post-trigger
Greater of 1 screen width or 50 µs
1 s to 500 s
Channel-to-channel deskew range
± 100 ns
Δ time accuracy (using cursors)
Same channel: ± (time base accuracy x reading) ± (0.0016 x screen width) ± 10 ps
Channel-to-channel: ± (time base accuracy x reading) ± (0.0016 x screen width) ± 15 ps
Main, zoom, roll, XY
On channels 1 and 2 only. Z blanking on ext trigger input, 1.4 V threshold
Bandwidth: maximum bandwidth. Phase error at 1 MHz: < 0.5 degree
Modes
XY
Horizontal system digital channels
Minimum detectable pulse width
Channel-to-channel skew
2 ns
2 ns (typical); 3 ns (maximum)
1. Denotes warranted specifications; All others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration