欢迎访问ic37.com |
会员登录 免费注册
发布采购

D/MSOX6000-AMG 参数 Datasheet PDF下载

D/MSOX6000-AMG图片预览
型号: D/MSOX6000-AMG
PDF下载: 下载PDF文件 查看货源
内容描述: [InfiniiVision 6000 X-Series Oscilloscopes]
分类和应用:
文件页数/大小: 45 页 / 3290 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号D/MSOX6000-AMG的Datasheet PDF文件第34页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第35页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第36页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第37页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第39页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第40页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第41页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第42页  
38 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet  
Performance Characteristics (Continued)  
Waveform math  
Number of math functions  
Arithmetic  
Four, displays all four simultaneously. Can be cascaded  
Add, subtract, multiply, divide, differentiate, integrate, FFT, Ax + B, squared, square root, absolute  
value, common logarithm, natural logarithm, exponential, base 10 exponential, low-pass filter,  
high-pass filter, averaged value, magnify, max hold, min hold, measurement trend, chart logic bus  
(timing or state), clock recovery  
Enhanced FFT  
Record size  
Window types  
Display  
Up to 1-Mpts resolution via precision mode  
Hanning, flat top, rectangular, Blackman-Harris, Bartlett  
Color grade or monochrome  
Waveforms  
Peak search  
FFT, max hold, min hold, average  
Max 11 peaks, threshold and excursion control  
Waveform analysis  
Mask/limit test (Option)  
Standard mask/limit test capability provides pass/fail comparison of a signal under test to a  
predefined mask template or automask template. Predefined mask templates or edits to an automask  
template can be made via a text editor. > 130,000 mask tests per second (waveform update rate)  
Provides a statistical view of a waveform or a measurement  
Any analog channels, math functions, reference waveforms, measurements  
Horizontal, vertical, or measurement  
Histogram  
Source  
Types  
Measurements  
Modes  
Hits, peak, max, min, peak to peak, mean, median, mode, bin width, standard deviation, 1~3 sigma  
All modes supported except zoom, ZY, and roll  
Color grade  
Provides a 3-dimensional view of waveform intensity  
Source  
Color themes  
Modes  
Any analog channels, math functions, reference waveforms, real-time eye  
Temperature and intensity  
All modes supported except zoom, ZY, and roll  
Jitter (Option)  
Measures the variance of a measurement over time  
Jitter measurement  
floor  
600 fs rms at 6 GHz sine wave (typical)  
Source  
Clock recovery  
Any analog channels, math functions, and reference waveforms  
Constant frequency, first-order phase lock loop (PLL), second-order PLL, explicit  
Data rate: Fully automatic, semi-automatic, manual  
Real-time eye (Option)  
Provides the color graded eye pattern analysis based on the recovered clock. Data bits are folded on  
top of each other per clock cycle to give a 3-dimensional view  
Any analog channels, math functions, and reference waveforms  
Constant frequency, first-order phase lock loop (PLL), second-order PLL, explicit  
Data rate: Fully automatic, semi-automatic, manual  
Source  
Clock recovery  
Color mode  
Color grade  
Measurements  
Eye height, eye width  
Precision mode  
Increase the analysis record length. Minimum: 100 kpts; maximum 1 Mpts  
 复制成功!