 浏览型号D/MSOX6000-AMG的Datasheet PDF文件第30页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第31页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第32页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第33页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第35页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第36页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第37页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第38页  
34 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet  
Performance Characteristics (Continued)  
Acquisition system  
1 GHz  
2.5 GHz  
4 GHz  
6 GHz  
Maximum analog channels sample rate  
Analog channels equivalent sample rate  
20 GSa/s half channel interleaved, 10 GSa/s all channel  
Not available  
400 GSa/s  
Maximum analog  
channels record length  
Maximum digital channels sample rate  
Maximum digital channels record length  
Acquisition mode  
≤ 2 GSa/s  
> 2 GSa/s  
4 Mpts half channel interleaved, 2 Mpts all channel  
1 Mpts half channel interleaved, 500 kpts all channel  
2 GSa/s half pods interleaved, 1 GSa/s all pods  
4 Mpts half pods interleaved, 2 Mpts all pods  
Default mode  
Normal  
Peak detect  
Analog channels: Capture glitches as narrow as 500 ps (half channel), 1 ns (all channel)  
Digital channels: Capture glitches as narrow as 500 ps (half pods), 1 ns (all pods)  
Selectable from 2, 4, 8, 16, 64, ... to 65,536  
Averaging  
High resolution  
Real-time boxcar averaging reduces random noise and effectively increases vertical resolution  
12 bits: ≥ 20 μs/div at 2 GSa/s or ≥ 50 μs/div at 1 GSa/s  
Segmented  
Segmented memory optimizes available memory for data streams that have long dead times  
between activity. Maximum segments = 1000. Re-arm time = As fast as 1 μs (minimum time  
between trigger events)  
Data acquisition mode  
Time mode  
Real time  
Max update rate  
Default mode (> 135,000 waveforms/sec)  
Enhanced real-time mode for the fastest waveform update rate of > 450,000 waveforms/sec. Up to  
2 GSa/s. Returns to a normal real-time mode at > 2 GSa/s  
Available with 6-GHz bandwidth license. The time base must be at 20 ns/div or faster. 2.5-ps fine  
interpolator resolution yields a maximum effective sample rate of 400 GSa/s  
Default mode  
Displays the waveform moving across the screen from right to left. Available at time bases 50 ms/div  
or slower  
Equivalent  
Normal  
Roll  
XY  
Shows the volts-versus-volts display. Time base can be set from 200 ns/div to 50 ms/div  
Trigger system  
Trigger sources  
Trigger modes  
Analog channel (1~4), digital channel (D0~D15), line, external, WaveGen (1, 2, or Mod (FM/FSK))  
Requires trigger event for oscilloscope to trigger  
Triggers automatically but not synchronized to the input in absence of trigger event  
Front panel button that triggers only once on a trigger event. Press Single button again for  
oscilloscope to find another trigger event, or press Run front panel button to trigger continuously in  
either Auto or Normal mode  
Normal  
Auto  
Single  
Force  
Front panel button forces a synchronous trigger  
Trigger coupling  
DC  
DC-coupled trigger  
AC  
AC-coupled trigger, cutoff frequency: < 10 Hz (internal); < 50 Hz (external)  
High-frequency reject, cutoff frequency ~ 50 kHz  
Low-frequency reject, cutoff frequency ~ 50 kHz  
Adds hysteresis to the trigger circuitry; selectable OFF or ON, decreases sensitivity 2x  
40 ns to 10.00 s  
HF reject  
LF reject  
Noise reject  
Trigger holdoff range  