欢迎访问ic37.com |
会员登录 免费注册
发布采购

D/MSOX6000-AMG 参数 Datasheet PDF下载

D/MSOX6000-AMG图片预览
型号: D/MSOX6000-AMG
PDF下载: 下载PDF文件 查看货源
内容描述: [InfiniiVision 6000 X-Series Oscilloscopes]
分类和应用:
文件页数/大小: 45 页 / 3290 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号D/MSOX6000-AMG的Datasheet PDF文件第37页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第38页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第39页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第40页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第42页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第43页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第44页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第45页  
41 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet  
Performance Characteristics (Continued)  
Dual-channel WaveGen — built-in function/arbitrary waveform generator (Specification are typical) (Continued)  
Amplitude  
Range: Minimum  
20 mVpp if |offset| ≤ 0.5 Vpp into Hi-Z 1  
10 mVpp if |offset| ≤ 0.5 Vpp into 50 Ω 1  
10 Vpp except, 9 Vpp if sinc or cardiac, 7.5 Vpp if Gaussian pulse into Hi-Z;  
5 Vpp/4.5 Vpp into 50 Ω  
Range: Maximum  
Resolution  
Accuracy  
Range  
100 μV or 3 digits, whichever is higher  
1.5% (frequency = 1 kHz)  
± 5 V into Hi-Z, except ± 4 V if sine wave, ± 2.5 V if sinc, cardiac, or Gaussian  
pulse into Hi-Z  
DC offset  
± 2.5 V into Hi-Z, except ± 2 V if sine wave, ± 1.25 V if sinc, cardiac, or  
Gaussian pulse into 50 Ω  
Resolution  
Larger of 250 uV or 3 digits  
Accuracy (waveform modes)  
Accuracy (DC mode)  
Impedance  
± 1.5% of offset setting ± 1% of amplitude ± 1 mV  
± 1.5% of offset setting ± 3 mV  
50 Ω typical  
Main output  
Isolation  
Protection  
Not available, main output BNC is grounded  
Overload automatically disables output  
Trigger output  
Trigger output available on trig-out BNC  
Quick action customization key  
Quick measure all  
Quick measure statistics reset  
Quick mask statistics reset  
Quick histogram statistics reset  
Quick print  
Displays a popup containing a snapshot of all the single waveform measurements  
Resets all measurement statistics and the measurement count  
Resets mask statistics and counters  
Zeros the histogram counters  
Print the current screen image  
Quick save  
Quick email  
Quick recall  
Saves the current setup, screen image, or data file as specified in the settings menu  
Emails the current setup, screen image, or data file as specified in the settings menu  
Recalls setup, mask, or reference waveform  
Quick freeze display  
Freezes the display without stopping running acquisitions or unfreezes the display if currently frozen.  
Waveform intensity preserved  
Quick trigger mode  
Quick clear display  
Toggles the trigger mode between auto and normal  
Clears the display  
Display characteristics  
Display  
Display mode  
Resolution  
Graticules  
12.1-inch capacitive multi-touch/gesture enabled color TFT LCD  
Zone/zoom/annotation mode and waveform placement mode  
800 (H) x 600 (V) pixel format (screen area)  
8 vertical divisions by 10 horizontal divisions with intensity controls  
YT and XY  
Format  
Maximum waveform update rate  
> 135,000 wfm/s (real time)  
> 450,000 wfm/s (real time max update rate)  
Off, infinite, variable persistence (100 ms to 60 s)  
256 intensity levels  
Persistence  
Intensity gradation  
1. Sinc, cardiac and Gaussian pulse: ± 1.25 V into Hi-Z; ± 625 mV into 50 Ω.  
 复制成功!