欢迎访问ic37.com |
会员登录 免费注册
发布采购

D/MSOX6000-AMG 参数 Datasheet PDF下载

D/MSOX6000-AMG图片预览
型号: D/MSOX6000-AMG
PDF下载: 下载PDF文件 查看货源
内容描述: [InfiniiVision 6000 X-Series Oscilloscopes]
分类和应用:
文件页数/大小: 45 页 / 3290 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号D/MSOX6000-AMG的Datasheet PDF文件第35页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第36页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第37页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第38页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第40页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第41页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第42页浏览型号D/MSOX6000-AMG的Datasheet PDF文件第43页  
39 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet  
Performance Characteristics (Continued)  
Precision counter/totalizer (Specifications are typical) (option)  
Counter  
Source  
Any analog channel or trigger qualified event  
Resolution  
10 digits (8 digits for trigger qualified event)  
Max frequency  
Trig-qual events  
Up to 3.2 GHz (4 GHz typical). With Hi-speed USB 2.0 decoding, 1 GHz (1.2 GHz typical)  
1/(trigger holdoff time) for trigger qualified events (max 25 MHz, minimum dead time of 40 ns)  
Frequency, period, totalize  
Measurement  
Totalizer  
Counter size  
Edge  
64-bit totalizing counter  
Rise or fall  
Gating  
Positive or negative level. Select from analog channels except the source  
Internal or external 10 MHz reference clock  
Time reference  
Integrated digital voltmeter (Specification are typical) (option)  
Source  
Analog channels only (1 ~ 4)  
Functions  
Resolution  
ACrms, DC, DCrms, frequency  
ACV/DCV: 3 digits  
Counter frequency: 5.5 digits  
Measuring rate  
Auto ranging  
Range meter  
100 times/second  
Automatic adjustment of vertical amplification to maximize the dynamic range of measurements  
Graphical display of most recent measurement, plus extreme over the previous 3 seconds  
Dual-channel WaveGen — built-in function/arbitrary waveform generator (Specification are typical) (Option)  
WaveGen outputs/  
Output modes  
Two (front-panel BNC connectors)  
Both waveform generator outputs can be frequency tracked, amplitude tracked, or completely tracked. Phase adjustable. 1  
A generator’s output can be inverted to create a differential signal  
Output modes: Normal (continuous) or single-shot (limited to arbitrary, sine, ramp, sine cardinal, exp rise/fall, cardiac,  
Gaussian pulse)  
Waveforms  
Modulation  
Sine, square, ramp, pulse, DC, noise, sine cardinal (sinc), exponential rise, exponential fall, cardiac, Gaussian pulse, and  
arbitrary  
Modulation is available on channel 1 only. Modulation is not available when tracking mode is enabled  
Modulation types: AM, FM, FSK  
Carrier waveforms: sine, ramp, sine cardinal, exponential rise, exponential fall, and Cardiac.C176  
Modulation source: Internal (no external modulation capability)  
AM:  
Modulation: Sine, square, ramp  
Modulation frequency: 1 Hz to 20 kHz  
Depth: 0 to 100%  
FM:  
Modulation: Sine, square, ramp  
Modulation frequency: 1 Hz to 20 kHz  
Minimum carrier frequency: 10 Hz  
Deviation: 1 Hz to carrier frequency or (2e12/carrier frequency), whichever is smaller  
FSK:  
Modulation: 50% duty cycle square wave  
FSK rate: 1 Hz to 20 kHz|  
Hop frequency: 2x FSK rate to 10 MHz  
1. Only the following combination of wave shapes can be frequency tracked or completely tracked:  
a. Sine, ramp, sine cardinal, cardiac, and Gaussian pulse.  
b. Square wave, and pulse.  
c. Exponential rise and exponential fall.  
d. Arbitrary.  
 复制成功!