26 | Keysight | CX3300 Series Device Current Waveform Analyzer - Data Sheet
Table 25. CX1102A other characteristics
Rise time (10% to 90%)
Input common mode impedance 1
Measurable over range
Burden voltage
0.35 / Bandwidth [MHz]
750 MΩ // 18 pF (Nominal)
10% of range
RIN * Measured current
±12 V
Maximum input voltage (common mode) 2
Peak voltage (DC + AC) limit
Range
Protection
1 A primary
200 mA primary
20 mA secondary
2 mA secondary
20 mA primary
2 mA primary
1.5 A fuse protected 3
50 mA fuse protected
Absolute maximum input current
200 μA secondary
20 μA secondary
1. Measured with CX1201A. Both inputs has this same input impedance. When using a CX1203A Sensor Head,
the minus (-) terminal is internally connected to the circuit common through a 10 MΩ resistor.
2. All current measurement ranges.
3. CX1203A with 50 Ω setting: 70 mA maximum. 125 mA fuse protected.
Table 26. CX1102A general information 1
Cable length
Dimension 2
Weight
Sensor cable: 1.5 m, Gnd lead: 16 cm
46.8 mm (W), 31.9 mm (H), 205.3 mm (D)
400 g
1 each Coaxial Termination Adapter Sensor Head (CX1203A)
1 each SMA(P) to BNC(J) 50 Ω Coaxial Adapter
1 each GND lead
Furnished accessories
1. Refer to mainframe’s “Environmental and general” part for other information.
2. CX1203A Sensor Head is included. Cable and adapter are not included.