30 | Keysight | CX3300 Series Device Current Waveform Analyzer - Data Sheet
CX1152A Digital Channel characteristics (mainframe: CX3324A only)
Warranted specifications are denoted by **, and all others are supplemental characteristics.
Specifications are valid after a 30-minute warm-up and 23 ± 5 ºC.
Table 35. CX1152A vertical system - digital channels
Input channels
8 channels
User-defined threshold range
Maximum input voltage
Threshold accuracy
Input dynamic range
Minimum input voltage swing
Input impedance
±25 V, 10 mV step
±40 V peak
±(150 mV + 3% of threshold setting)
±25 V
500 mV peak-to-peak
10 MΩ ±2% with approximately 8 pF in parallel
Channel-to-channel skew
Resolution
4 ns
1 bit
1. 50 Ω input impedance
2. CX1152A is required to enable the input digital channels.
Table 36. CX1152A general information
Cable length
Dimension 1
Weight
Digital channel cable: 1.15 m, probe lead: 28.5 cm
68.1 mm (W), 18.5 mm (H), 103.0 mm (D)
130 g
5 probe ground leads
10 grabbers
Furnished accessories
1 each BNC-probe tip adapter
1. Dimension of pod. Leads and cables are not included.