23 | Keysight | CX3300 Series Device Current Waveform Analyzer - Data Sheet  
CX1100 Series Current Sensors performance characteristics  
Warranted specifications are denoted by **, and all others are supplemental characteristics. Specifications are valid after a 30-minute warm-up and 23 ±5 ºC.  
All these characteristics are defined by 14-bit acquisition resolution unless otherwise stated.  
About the measurement accuracy  
Measurement accuracy can be affected by RF electro-magnetic field having the strengths greater than 3 V/m in the frequency range of 80 MHz to 2 GHz or  
1 V/m in the frequency range of 2 GHz to 2.7 GHz. The extent of this effect depends upon how the instrument is positioned and shielded.  
Table 19. CX1101A characteristics overview  
Noise (rms) 3  
Maximum bandwidth  
(-3 dB) 4  
2
Range  
RIN  
10 A  
1 A  
15 mΩ (typ)  
10 mA  
2 mA  
3 MHz 5  
100 MHz  
100 MHz  
100 MHz  
200 mA  
20 mA  
2 mA  
0.2 mA  
20 μA  
410 mΩ (typ)  
550 mΩ (max)  
Current measurement 1  
3 μA  
400 nA  
40 nA  
100 MHz  
25 kHz  
25 kHz  
200 μA  
20 μA  
50 Ω (typ)  
77 Ω (max)  
1. Sensor Head used to measure the characteristics: CX1206 for 10 A range and CX1203A for all other ranges.  
2. The slide switch of CX1203A to be set to “0 Ω”.  
3. 20 MHz noise bandwidth measured with mainframe  
4. Standalone bandwidth. The effective measurement bandwidth when connected to mainframe can be estimated by the following equation.  
1
BWeffective  
=
1
1
BWmainframe  
BWsensor  
2
2
+
5. -4 dB bandwidth  
Table 20. CX1101A DC measurement accuracy 1  
Range  
Standalone  
23 ±5 ºC  
With mainframe  
TUSERCAL ±3 ºC,  
24 hrs 2  
23 ±5 ºC  
10 A  
± (5% + 5%)  
± (2% + 2%) **  
± (2% + 2%) **  
± (2% + 2%) **  
± (2% + N/A) **  
± (2% + 2%) **  
± (2% + N/A) **  
± (5.7% + 5.9%)  
± (2.7% + 2.9%) **  
± (2.7% + 2.9%) **  
± (2.7% + 2.9%) **  
± (2.7% + N/A) **  
± (2.7% + 2.9%) **  
± (2.7% + N/A) **  
N/A  
1 A  
± (1.8% + 0.4%)  
± (0.7% + 0.4%)  
± (0.6% + 0.3%)  
± (0.7% + 1.1%)  
± (0.7% + 0.3%)  
± (0.7% + 1.1%)  
200 mA  
Gain [% of readings] +  
Offset [% of range]  
20 mA  
2 mA  
200 μA  
20 μA  
1. Accuracy is defined at VCM = 0 V (zero Common Mode input voltage at either +IIN or -IIN). Add 0.7% typical to Offset error for Vcm up to 40 V.  
The “readings” means measured data by mainframe. DC measurement condition: 20 ms averaged.  
2 After executing the User Calibration with mainframe. Supplemental characteristics.  