16 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet
Device Capacitance Selector (N1272A) specification
The N1272A simplifies 2 and 3 terminal device capacitance measurements by automatically creating the correct configu-
ration of test resources (including adding any needed DC blocking capacitors and AC blocking resistors) for a specified
capacitance measurement. To measure packaged device capacitance the N1273A Capacitance Test Fixture is also neces-
sary. However, the N1272A can be used directly with a probe station to measure on-wafer device capacitances.
DC bias characteristics
100 kΩ at SMU bias output resistance
Voltage drop compensation function is available.
Bypass capacitance in the capacitance selector
Capacitance
1 μF
Withstand voltage
±3000 V
Drain to source terminal
Gate to source terminal
1 μF
±100 V
Measurement accuracy for 2-terminal device (Supplemental characteristics)
The accuracy of the supplemental characteristics is defined at the output terminals of the TO socket adapter in the
N1273A Capacitance Test Fixture when the N1272A is connected to B1505A with the 1.5 m CMU cable and the N1273A
system cable.
Measurement accuracy for 2-terminal device
1.E-06
Condition
AC level: 30 mV rms
Dx < = 0.1
1.E-07
(Dx: Measurement value of D)
1.E-08
1.E-09
1.E-10
Error 3%
1.E-11
Error 5%
Error 10%
1.E-12
Error 20%
1.E-13
1.E-14
1.E-15
1.E+03
1.E+04
1.E+05
1.E+06
Frequency (Hz)
Output terminals for 2-terminal device
Collector/drain
Emitter/source
Base/gate
High
High
Low
High
Low
Low