CDB5521/22/23/24/28  
down menu above the register decode box to  
select between the different registers.  
2.4 Setup Window Overview  
The Setup Window (Figure 6) allows the user to  
read and write to the internal register of the con-  
verter in either binary or hexadecimal, and acquire  
real-time conversions. It has quick access control  
icons that quickly reset the converter, reset the con-  
verter’s serial port, or self-calibrate the converter’s  
offset and gain. The following are controls and in-  
dicators associated with this window.  
Hexadecimal Icons  
Nine input/display icons that allow a user to  
set/clear the 24 bits in the configuration, or  
channel setup registers via 6 hexadecimal nib-  
bles. If the upper nibbles in the registers are ze-  
ro’s, the leading zero nibbles need not be  
entered.  
Acquire Data  
Num To Average  
This is a control icon. When pressed, the PC  
transmits the collect single conversion command  
to the microcontroller. The microcontroller in  
turn collects a conversion from the ADC and re-  
turns it to the PC. The PC stores the conversion  
and collects additional conversions to form a set.  
From the sample set collected, the high, the low,  
peak-to-peak, average, and standard deviation,  
are computed (the size of the data set is set by the  
Num To Average input) and then the display  
icons are updated. This process continues until  
the STOP button is pressed, or until another win-  
dow is selected.  
Input icon that sets the size of the data conver-  
sion set referred to after the Acquire Icon is ac-  
tivated.  
Reinitialize Port  
This is a control icon. When pressed, 128 logic  
1’s followed by a logic ‘0’ are sent to the  
ADC’s serial port to reset its port. It does not re-  
set the RS-232 link.  
Reset A/D  
This is a control icon. When pressed, the micro-  
controller sends the appropriate commands to  
return the converter to its initial default state.  
Note: The quick access control icons are disabled  
once Acquire is selected. This eliminates  
potential hardware conflicts.  
Stop  
Stops the collection of conversion data.  
Binary Icons  
Update Icons  
Input icons array to set/clear the 24 individual  
bits in the configuration or channel-setup regis-  
ters. The respective registers bit is set/cleared  
as soon as the icon is clicked.  
This is a control icon. When pressed the config-  
uration and channel -setup registers contents  
are acquired. Then, the configuration text box  
and the register content icons are updated.  
Channel  
Data FIFO Window  
Selects the Setup that will be accessed to per-  
form conversions when Acquire Data is activat-  
ed.  
This button opens the Data FIFO Window  
when pressed.  
2.5 Data FIFO Window Overview  
Register Decode Box  
The following describes the controls available in  
the Data FIFO Window (Figure 7).  
Text display box that displays the decoded  
meaning of each bit in the configuration regis-  
ter and the channel setup registers. Use the pull-  
DS317DB2  
15  
 复制成功!