CDB5521/22/23/24/28  
Update Icons  
2.9 Calibration Window Overview  
The following controls and indicators are associat-  
ed with the Calibration Menu (Figure 9).  
This is a control icon. When pressed the offset  
and gain registers are read. Then, the register  
content icons are updated.  
GAIN DECODE  
2.10 Trouble Shooting the Evaluation Board  
Eight display boxes that displays the decoded  
meaning of each gain register.  
This section describes special test modes incorporated  
in the microcontroller software to diagnose hardware  
problems with the evaluation board.  
Gain Hexadecimal Icons  
Eight input/display icons that allow a user to  
set/clear the 24 bits in the eight gain registers  
via 6 hexadecimal nibbles. If the upper nibbles  
in the registers are zero’s, then leading zero  
nibbles need to be entered.  
Note: To enter these modes, set the test switches to the  
appropriate position and reset the evaluation  
board. To re-enter thenormal operationmode, set  
the switches back to binary zero and reset the  
board again.  
Test Mode 0, Normal Mode  
Offset Hexadecimal Icons  
This is the default mode of operation. To enter  
this mode, set the test switches to 000 and reset  
the board. The evaluation board allows normal  
read/writes to the ADC’s registers. All the  
LED’s toggle on then off after reset, and then  
only when communicating with the PC.  
Eight input/display icons that allow a user to  
set/clear the 24 bits in the eight offset registers  
via 6 hexadecimal nibbles. If the upper nibbles  
in the registers are zero’s, then leading zero  
nibbles need to be entered.  
Self-Gain  
Test Mode 1, Loop Back Test  
Used to perform a self-gain calibration using  
the chosen Setup.  
This test mode checks the microcontroller’s on-  
chip UART. To enter this mode, set test switch-  
es to 001, set HDR7 for loop back, and then re-  
set the board. If the communication works, all  
the LED's toggle. Otherwise, only 1/2 of the  
LED’s toggle to indicate a communication  
problem.  
Self-Offset  
Used to perform a self-offset calibration using  
the chosen Setup.  
Shift Gain Register  
Sixteen input icons used to shift the contents on  
the gain registers either 1 bit left or right. Once  
shifted the data at the respective gain registers  
ends is lost.  
Test Mode 2, Read/Write to ADC  
This test mode tests the microcontroller’s abil-  
ity to read and write to the ADC. To enter this  
mode, set the switches to 010 and reset the  
board. In this test mode, the ADC’s configura-  
tion, offset, and gain registers are written to and  
then read from. If the correct data is read back,  
all the LED's toggle. Otherwise, only half of  
them toggle to indicate an error.  
System-Gain  
Used to perform a system-gain calibration us-  
ing the chosen Setup.  
System-Offset  
Used to perform a system-offset calibration us-  
ing the chosen Setup.  
DS317DB2  
19  