13  
n Main unit Specifications  
Additional features  
Measurement parameter printing, cursor measurement, scaling,  
current clamp setting, comment input, screen hard copy, list/gauge,  
start condition hold, auto setup, auto save, remote control (start/stop/  
print control), auto range, over-range indication, VIEW function, key  
lock, level monitor, vernier function, offset cancel, event marker  
input, waveform search function, report printing  
General  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,  
approx. 290 g (10.2 oz) Accessories: None  
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 30 minutes of warm-  
up time and zero-adjust; accuracy guaranteed for 1 year)  
DC/RMS UNIT 8959  
For the 8860 series only  
n
Options specifications (sold separately)  
Number of channels: 2, for voltage measurementl, DC/RMS selectable  
Measurement functions  
Isolated BNC connector (input impedance 1MΩ, input capacitance 30pF), Max. rated  
voltage to earth: 370V AC, DC (with input isolated from the unit, the maximum voltage that  
can be applied between input channel and chassis and between input channels without damage)  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,  
approx. 290 g (10.2 oz) Accessories: None  
Input connectors  
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 30 minutes of warm-up  
time and zero-adjust; accuracy guaranteed for 1 year)  
ANALOG UNIT 8956  
Measurement functions  
5mV to 20V/DIV, 12 ranges, full scale: 20DIV, AC voltage for  
possible measurement/display using the memory function: 280V  
rms, low-pass filter: 5Hz/500Hz/5kHz/100kHz  
Measurement  
range  
Number of channels: 2, for voltage measurement  
Isolated BNC connector (input impedance 1MΩ, input capacitance 40pF), Max. rated  
voltage to earth: 300V AC, DC (with input isolated from the unit, the maximum voltage that  
can be applied between input channel and chassis and between input channels without damage)  
1/100 of measurement range (using 12-bit A/D conversion; installed in 8860 series)  
Measurement resolution  
Highest sampling rate  
Input connectors  
1MS/s (simultaneous sampling in 2 channels)  
DC amplitude: ±0.4% of full scale (with filter 5Hz), zero position: ±0.1%  
of full scale (with filter 5Hz, after zero adjustment)  
5mV to 20V/DIV, 12 ranges, full scale: 20 DIV, AC voltage for possible  
measurement/display using the memory function: 280V rms, low-pass  
filter: 5Hz/500Hz/5kHz/1MHz  
Accuracy  
Measurement  
range  
RMS amplitude accuracy: ±1% of full scale (DC, 20Hz to 1kHz), ±3% of full scale (1kHz to  
100kHz), response time: SLOW 5s (rise time from 0 to 90% of full scale), MID 800ms (rise time  
from 0 to 90% of full scale), FAST 100ms (rise time from 0 to 90% of full scale), crest factor: 2  
RMS  
measurement  
1/100 of measurement range (using 12-bit A/D conversion; installed in 8860 series)  
Measurement resolution  
Highest sampling rate  
20MS/s (simultaneous sampling in 2 channels)  
DC to 400kHz ±3dB, with AC coupling: 7Hz to 400kHz ±3dB  
DC, GND, AC  
Frequency characteristics  
Input coupling  
DC amplitude: ±0.4% of full scale (with filter 5Hz)  
Zero position: ±0.1% of full scale (with filter 5Hz, after zero adjustment)  
Accuracy  
400V DC (the maximum voltage that can be applied across input pins without damage)  
Max. allowable input  
DC to 10MHz ±3dB, with AC coupling: 7Hz to 10MHz ±3dB  
DC, GND, AC  
Frequency characteristics  
Input coupling  
400V DC (the maximum voltage that can be applied across input pins without damage)  
Max. allowable input  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,  
approx. 290 g (10.2 oz) Accessories: Conversion cable × 2, cable length 50cm (19.69in)  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,  
approx. 310 g (10.9 oz) Accessories: None  
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up  
time and auto-balance; accuracy guaranteed for 1 year)  
STRAIN UNIT 8960  
Measurement functions  
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 30 minutes of warm-  
Number of channels: 2, for distortion measurement (electronic auto-balancing, balance adjustment range within ±10000 με)  
HIGH-RESOLUTION UNIT 8957  
up time and zero-adjust; accuracy guaranteed for 1 year)  
Via conversion cable, TAJIMI PRC03-12A10-7M10.5, Max. rated voltage to  
earth: 33Vrms or 70V DC (with input isolated from the unit, the maximum voltage that can  
be applied between input channel and chassis and between input channels without damage)  
Number of channels: 2, for voltage measurement  
Measurement functions  
Input connectors  
Isolated BNC connector (input impedance 1MΩ, input capacitance 40pF), Max. rated  
Input connectors  
voltage to earth: 300V AC, DC (with input isolated from the unit, the maximum voltage that  
can be applied between input channel and chassis and between input channels without damage)  
Strain gauge converter, bridge impedance: 120Ω to 1kΩ (bridge voltage 2V),  
350Ω to 1kΩ (bridge voltage 5V, 10V), bridge voltage 2, 5, 10 ±0.05V  
Suitable  
transducer  
5mV to 20V/DIV, 12 ranges, full scale: 20DIV, AC voltage for  
possible measurement/display using the memory function: 280V  
rms, low-pass filter: 5Hz/50Hz/500Hz/5kHz/50kHz  
Measurement  
range  
20με to 1000με/DIV, 6 ranges, full scale: 20DIV, low-pass filter:  
5Hz/10Hz/100Hz/1kHz  
Measurement  
range  
Integrated filter for suppressing aliasing distortion caused by FFT  
processing (automatic cutoff frequency setting/OFF)  
Integrated filter for suppressing aliasing distortion caused by  
FFT processing (automatic cutoff frequency setting/OFF)  
Anti-aliasing filter  
Anti-aliasing filter  
1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860 series)  
Measurement resolution  
Highest sampling rate  
1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860 series)  
Measurement resolution  
Highest sampling rate  
200kS/s (2-channel simultaneous sampling)  
2MS/s (simultaneous sampling in 2 channels)  
DC amplitude: ±(0.4% of full scale +2με), zero position: ±(0.1%  
of full scale +2με) (at 5Hz filter ON)  
DC amplitude: ±0.2% of full scale (with filter 5Hz)  
Zero position: ±0.1% of full scale (with filter 5Hz, after zero adjustment)  
Accuracy After auto-balancing  
Accuracy  
DC to 20kHz +1/3dB  
Frequency characteristics  
Max. allowable input  
DC to 200kHz ±3dB, with AC coupling: 7Hz to 200kHz ±3dB  
DC, GND, AC  
Frequency characteristics  
Input coupling  
10V DC (the maximum voltage that can be applied across input pins without damage)  
* Available from main unit 8860/8861 version 1.06  
400V DC (the maximum voltage that can be applied across input pins without damage)  
Max. allowable input  
Dimensions and mass: approx. 170 (6.69in) W × 19.8 (0.78in) H  
× 148.5 (5.85in) D mm, approx. 310 g (10.9 oz)  
Accessories: CONNECTION CORD 9ꢀ4ꢀ× 2, GRABBER CLIP 9ꢀ4ꢁ× 2  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 183 (7.20in) D mm,  
approx. 385 g (13.6 oz) Accessories: Flathead screwdriver × 1, short bar × 2  
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 1 hour of warm-up  
(Accuracy at 23 ±5°C/73 ±9°F, 30 to 80 % rh after 30 minutes of warm-up  
time and zero-adjust; accuracy guaranteed for 1 year)  
16ch SCANNER UNIT 8958  
HIGH VOLTAGE UNIT 8961  
time and adjustment; accuracy guaranteed for 1 year)  
Number of channels: 16, for voltage measurement/temperature measurement with thermocouple  
Measurement functions  
Number of channels: 2, for voltage measurementl, DC/RMS selectable  
Measurement functions  
Safety Banana Connector (input impedance 10MΩ, input capacitance 5pF), Max. rated  
voltage to earth: 1000V AC, DC CAT II, 600V AC, DC CAT III (the maximum voltage  
that can be applied between input channel and chassis and between input channels without damage)  
Voltage input/Thermocouple input: screw-type terminal strip, recommended  
wire diameter *1, detachable terminal block (with cover) *1 Recommended cable, single-  
wire: 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16  
Input impedance: 1MΩ, 850kΩ with line fault detection ON, Max. rated  
voltage to earth: 33Vrms or 70V DC (with input isolated from the unit, the maximum  
Input connectors  
Input connectors  
1V to 50V/DIV, 6 ranges, full scale: 20 DIV, AC voltage for possible  
measurement/display using the memory function: 700V rms, low-pass  
filter: 5Hz/50Hz/500Hz/5kHz  
Measurement  
range  
voltage that can be applied between input channel and chassis and between input channels  
without damage)  
1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860 series)  
Measurement resolution  
Highest sampling rate  
5m, 50m, 500m, 2V/DIV, 4 ranges, full scale: 20DIV, measurement range: ±100%  
of full scale, digital filter: 10Hz/50Hz/60Hz, measurement resolution 1/1600 of  
measurement range (using 16-bit A/D conversion; installed in 8860 series)  
Voltage  
measurement  
range  
2MS/s (simultaneous sampling in 2 channels)  
DC amplitude: ±0.25% of full scale (with filter 5Hz)  
Zero position: ±0.15% of full scale (with filter 5Hz, after zero adjustment)  
Accuracy  
Temperature  
measurement range  
(Upper and lower limit values  
depend on measurement input  
range of sensor)  
10°C/DIV (-100°C/ to +200°C), 50°C/DIV (-200°C/ to +1000°C), 100°C/DIV  
(-200°C/ to +2000°C), 3 ranges, full scale: 20DIV, digital filter: 10Hz/50Hz/  
60Hz, measurement resolution 1/1000 of measurement range (using 16-bit  
A/D conversion; installed in 8860 series)  
RMS  
measurement  
RMS amplitude accuracy: ±1% of full scale (DC, 40Hz to 1kHz sin waveform),  
±3% of full scale (1kHz to 10kHz sin waveform), crest factor: 2  
DC to 100kHz ±3dB  
Frequency characteristics  
Input coupling  
K: -200 to 1350°C, J: -200 to 1200°C, E: -200 to 1000°C, T: -200 to  
400°C, N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C, B: 400 to  
1800°C, W (WRe5-26): 0 to 2000°C, reference junction compensation:  
internal/ external (switchable), line fault detection ON/OFF switchable  
Thermocouple  
range  
(JIS C 1602-1995)  
(ASTM E-988-96)  
DC, GND  
1000V DC (the maximum voltage that can be applied across input pins without damage)  
Up to four units settable for the 8860-50, or the 8861-50 one main unit.  
Max. allowable input  
Number of modules  
50ms/all channels (digital filter OFF), 300ms/all channels (digital  
filter 50Hz/60Hz), 1.4 s/all channels (digital filter 10Hz)  
Data refresh rate  
Voltage: ±0.2% of full scale, thermocouple (K, J, E, T, N): ±0.05%  
of full scale ±1°C, (R, S, B, W): ±0.05% of full scale ±2°C (400°C  
or more), ±0.05% of full scale ±3.5°C (less than 400°C), reference  
junction compensation accuracy: ±1°C (added to measurement accuracy  
with internal reference junction compensation)  
Accuracy  
40V DC (the maximum voltage that can be applied across input pins without damage)  
Max. allowable input