欢迎访问ic37.com |
会员登录 免费注册
发布采购

8861-50 参数 Datasheet PDF下载

8861-50图片预览
型号: 8861-50
PDF下载: 下载PDF文件 查看货源
内容描述: 存储记录仪 [MEMORY HiCORDER]
分类和应用: 存储
文件页数/大小: 16 页 / 2428 K
品牌: ASM-SENSOR [ ASM GMBH ]
 浏览型号8861-50的Datasheet PDF文件第8页浏览型号8861-50的Datasheet PDF文件第9页浏览型号8861-50的Datasheet PDF文件第10页浏览型号8861-50的Datasheet PDF文件第11页浏览型号8861-50的Datasheet PDF文件第13页浏览型号8861-50的Datasheet PDF文件第14页浏览型号8861-50的Datasheet PDF文件第15页浏览型号8861-50的Datasheet PDF文件第16页  
12  
n Main unit Specifications  
Recorder functions  
10 ms to 200 ms *8/division, 500 ms to 1 hour/div with 19 ranges, time  
axis resolution 100 points/division, time axis zoom: ×2 to ×4 in 2 stages,  
compression: 1/2 to 1/20,000 in 13 stages  
At recoding length “continuous”, time axis 20 ms/div to 1 hour/div  
With scanner module 8958, time axis 50ms/div to 1 hour/div  
6
Print/display section  
Display  
* Printer functions are available when optional printer unit is installed  
10.4 inch TFT color LCD (SVGA, 800 × 600 dots)  
RECORDING PAPER 9ꢀꢁ1: 216 mm (8.50 in) × 30 m (98.43 ft),  
thermal paper roll (when using A4-size the printer unit 8995)  
RECORDING PAPER 9ꢀꢁ4: 112 mm (4.41 in) × 18 m (59.06 ft),  
thermal paper roll (when using A6-size the printer unit 8995-01)  
Time axis  
6 Recording paper  
*
8: Virtual record function: At 10 ms - 200 ms/division, printing in real time is not possible, but  
*
waveform data are stored in memory and can be monitored on screen. Data are stored for 5,000  
divisions before the end of measurement. At recording length settings other than "Continuous", the  
printer can be used simultaneously, for follow-up printing of waveforms.  
RECORDING PAPER 9ꢀꢁ1: 200 mm (7.87 in) , full scale 20 divisions, 1  
division = 10 mm (0.39 in) (when using A4-size the printer unit 8995)  
RECORDING PAPER 9ꢀꢁ4: 100 mm (3.94 in) , full scale 10 divisions, 1  
division = 10 mm (0.39 in) (when using A6-size the printer unit 8995-01)  
100 ns to 1 sec in 8 stages (selectable in 1/100 of time axis range)  
Sampling rate  
6 Recording width  
*
32 MW memory: free setting in 1-division steps (max. 5,000 div), continuous *9, up  
to 1,000 divisions use with the scanner module 8958  
10 lines/mm (when using A4-size the printer unit 8995 ), 8 lines/mm  
(when using A6-size the printer unit 8995-01)  
* 20 lines/mm with "smoothed printing" memory function (when using A4-size the printer unit 8995)  
128 MW memory: free setting in 1-division steps (max. 20,000 div), continuous *9,  
up to 5,000 divisions use with the scanner module 8958  
6
*
*
Paper feed density  
512 MW memory: free setting in 1-division steps (max. 80,000 div), continuous *9,  
up to 20,000 divisions use with the scanner module 8958  
6 Recording speed Max. 25 mm (0.98 in)/sec  
Recording length  
1 GW memory: free setting in 1-division steps (max. 160,000 div), continuous *9,  
up to 40,000 divisions use with the scanner module 8958  
Trigger functions  
*9 At time axis 10 ms to 200 ms/division and printer ON, Continuous setting cannot be selected. At use  
with the Printer Unit 8995-01 and numerical print ON, time axis 10 ms to 1 sec/division is disabled  
Note: Memory of the 8861-50 is twice than shown above, but recording length is the same.  
Turn on/off independently for each trigger source of analog/logicA– D,  
external trigger (a rise of 2.5V or terminal short circuit); timer trigger,  
inter-sourceAND/OR, forced trigger,  
standard mode (trigger source to all analog channels settable), extend  
mode (multiple analog sources to a single analog channel settable, up  
to 8 for 8860-50, and up to 8 on channels/units 1 – 4, and up to 8 on  
channels/units 5 – 8 for 8861-50 settable)  
Store data for most recent 5,000 *10 divisions, or up to 160,000 div in  
Trigger sources  
memory. Backward scrolling and re-printing available.  
Waveform memory  
Screen and printing  
10  
*
Depending on the amount of installed memory, Memory of 8861-50 or 8861 is twice that  
of 8860-50 or 8860, but recording length is the same.  
Split screen (1 to 8), sheet display (max. 32 channels per sheet), logging  
(print/display measurement data as digital values), voltage axis zoom (×2 to ×  
100), compression (×1/2 to ×1/10), variable display  
Level: Triggering occurs when preset voltage level is crossed  
(upwards or downwards).  
Window: Triggering occurs when window defined by upper and  
lower limit is entered or exited.  
Period: Rising edge or falling edge cycle of preset voltage value  
is monitored and triggering occurs when defined cycle range is  
exceeded.  
Glitch: Triggering occurs when pulse width from rising or falling  
edge of preset voltage value is underrun.  
Slope: Triggering occurs when preset change degree (slope) is  
exceeded or underrun.  
Voltage drop: Triggering occurs when voltage drops below peak  
voltage setting (for 50/60 Hz AC power lines only).  
Event setting: Event count is performed for each source, and  
triggering occurs when a preset count is exceeded.  
REC & MEM function  
100 ms to 1 hour/division, 16 ranges, time axis resolution 100 points/  
division, sampling rate: same as sampling rate for MEM function.  
Recording data of the scanner module 8958 in REC side.  
Trigger types  
(analog)  
Time axis  
(REC)  
10 μs to 5 min/division, 25 ranges, time axis resolution 100 points/  
division, sampling rate: 1/100 of time axis  
Time axis  
(MEM)  
REC: 25 to 2,000 *11 divisions, or up to 80,000 div *11, continuous  
MEM: 25 to 5,000 *11 divisions, or up to 160,000 div *11  
Recording length  
11  
*
Depends on installed memory 32 MW to 1GW (free setting in 1-division steps also possible)  
The last 2,500 *11 divisions, or up to 80,000 div *11 are saved to memory  
for scroll-back and re-print.  
Limited according to whether 16-Ch Scanner Module 8958 is installed.  
0.1% of full scale (full scale = 20 divisions)  
Level setting resolution  
Waveform Memory  
1, 0, 01, ×, pattern setting, AND/OR setting for groups of 4 channels,  
level or edge detect selectable (01: changing to any value activates trigger)  
(REC)  
Trigger types  
(logic)  
Toggle REC/MEM waveform display, simultaneous display of REC/  
MEM waveform with split screen, split screen (1 to 8), sheet display  
(max. 32 channels per sheet), logging (print/display measurement data as  
digital values), zoom (with MEM), variable display  
OFF, setting range 0.1 to 10.0 divisions in 0.1 division steps (MEM, REC  
& MEM function), ON (10 ms)/OFF (REC function)  
Trigger filter  
Screen and  
printing  
(analog/logic)  
Pre-trigger function to capture pre- and post-trigger waveform,  
trigger output (active Low with terminal block and open collector  
5 voltage output). Level display while waiting for trigger, Start/  
stop trigger conditions independently selectable  
Other functions  
Divided use of memory space (up to 1024 divisions), sequential  
save, block search  
Memory divide  
FFT function  
Memory functions  
Storage waveform, linear spectrum, RMS spectrum, power  
spectrum, power spectrum density, cross power spectrum, power  
spectrum density (LPC), auto-correlation function, histogram,  
transfer function, cross-correlation function, phase spectrum,  
impulse response, coherence function, octave analysis  
5 μs to 5 min/division, 26 ranges or external sampling, time axis  
resolution 100 points/division, time axis zoom: ×2 to ×10 in 3  
stages, compression: 1/2 to 1/500,000 in 17 stages  
Time axis  
Analysis mode  
Fixed: 1/100 of time axis range, Variable: external sampling  
Sampling period can be used to set time axis  
Two different sampling rate settings are possible  
Sampling rate  
1-channel FFT, 2-channel FFT in selected channels (up to 16 analysis  
functions)  
Analysis channels  
32 MW memory : free setting in 1-division steps (max. 320,000 div *7) or  
built-in presets of 25 to 200,000 divisions *7  
133 mHz to 8 MHz, resolution 1/400, 1/800, 1/2000, 1/4000, 1/8000  
1000, 2000, 5000, 10000, 20000 points  
Frequency range  
No.of sampling points  
128 MW memory : free setting in 1-division steps (max. 1,280,000 div *7) or  
built-in presets of 25 to 1,000,000 divisions *7  
Selected from: Newly loaded data / MEM function waveform  
data / MEM waveform of REC & MEM function  
Analysis data  
512 MW memory : free setting in 1-division steps (max. 5,120,000 div *7) or  
built-in presets of 25 to 5,000,000 divisions *7  
Recording length  
Pre-trigger  
Rectangular, Hanning, Exponential, Hamming, Blackman,  
Blackman-Harris, Flat-top  
1 GW memory : free setting in 1-division steps (max. 10,240,000 div *7) or  
Window functions  
built-in presets of 25 to 10,000,000 divisions *7  
Maximum recording length or built-in preset length when using 1 channel (8860-50) or 2 channels  
(8861-50). Memory of 8861-50 is twice that of 8860-50, but recording length is the same.  
7
Split screen (1/2/4), Nyquist, logging (print/display measurement data as  
digital values), frequency axis zoom and left/right scrolling  
Screen and  
printing  
*
Record data from before the trigger point, -100 to +100% of  
recording length (free setting in 1% steps)  
Time axis / frequency axis simple averaging, exponential  
averaging, peak hold, (free settting 2 times to 10,000 times)  
Averaging  
Split screen (1 to 16), X-Y screen (1, 4 screens, max. 16 combined), sheet  
display (max. 32 channels per sheet), logging (print/display measurement data as  
digital values), voltage axis zoom (×2 to ×100), compression (×1/2 to ×1/10),  
overlay, zoom, variable display, vernier display  
Real-time save function  
Screen and  
printing  
10 ms to 1 hour/division, 19 ranges, time axis resolution 100 points/div,  
Time axis  
(Whole waveform data)  
sampling speed: same as sampling rate for “Measurement Waveform”  
Divided use of memory space (up to 4096 divisions), sequential  
save, block serch  
Time axis  
100 μs to 5 min/division, 22 ranges (limited depending on store target and number of  
channels), time axis resolution 100 points/div, sampling rate: 1/100 of time axis  
Memory splitting  
Waveform calculation  
(Measurement waveform  
data: sampling data)  
Four arithmetic operations, absolute value, exponentiation, common logarithm,  
square root, moving average, differentiation once and twice, integration once and  
twice, parallel displacement along the time axis, trigonometric functions (sin,  
cos, tan, arc-sin, arc-cos, arc-tan), Any of 16 calculation types can be applied to  
recording lengths of up to 1/4 of memory capacity  
HDD, PC via LAN, PC Card (use only HIOKI’s card)  
Save to  
Depending on available space on storage media / file system / number  
of channels / REC time axis, Selectable in division steps up to  
maximum recording length  
Recording length  
(Numerical calculation by specifying calculation area with cursors A and B, numerical calculation judgment,  
automatic saving of numerical calculation results, saving of any existing numerical calculation results)  
During measurement: Whole wave, after measurement: toggle  
Whole/Measurement waveform display, simultaneous display of Whole/  
Measurement waveform with split screen, split screen (1 to 8), 16 split  
(A4-size printer only), sheet display (max. 32 channels per sheet), logging (print/  
display measurement data as digital values), zoom, variable display  
Screen and  
printing  
Average value, effective (rms) value, peak to peak value, maximum value,  
time to maximum value, minimum value, time to minimum value, period,  
frequency rise time, fall time, area value, X-Y area value, standard deviation,  
time to level, pulse width, duty ratio, pulse count, Up to 16 items can be  
selected.  
Numerical calculation  
Averaging  
Data can be analyzed in MEM function or FFT function  
Memory transfer  
Detection of trigger criteria, time, event markers and peak value  
Up to 1,000 event markers can be input during and after measurement  
Waveform serch  
function  
Cumulative average, Exponential average (select 2 to 10,000 data objects  
to be averaged)