Agilent 16950B and 16951B Specifications and Characteristics
Timing (Asynchronous) Analysis Mode
Trigger sequence level branching
Trigger position
Conventional Timing
Transitional Timing
Arbitrary 4-way if/then/else
Start, center, end, or user-defined
Arbitrary 4-way if/then/else
Start, center, end, or user-defined
Trigger resources
16 patterns evaluated as =, =/, >, ≥, <, ≤
14 double-bounded ranges evaluated as
in range, not in range
15 patterns evaluated as =, =/, >, ≥, <, ≤
14 double-bounded ranges evaluated as
in range, not in range
3 edge/glitch
3 edge/glitch
2 timers per module
2 timers per module
2 global counters
2 global counters
1 occurrence counter per sequence level
4 flags
1 occurrence counter per sequence level
4 flags
Trigger resource conditions
Trigger actions
Arbitrary Boolean combinations
Arbitrary Boolean combinations
Go To
Go To
Trigger, send e-mail, and fill memory
Trigger and Go To
Trigger, send e-mail, and fill memory
Trigger and Go To
Turn on/off default storing
Timer start/stop/pause/resume
Global counter increment/decrement/reset
Occurrence counter reset
Flag set/clear
Turn on/off default storing
Timer start/stop/pause/resume
Global counter increment/decrement/reset
Occurrence counter reset
Flag set/clear
Maximum global counter
Maximum occurrence counter
Maximum pattern/range width
Maximum pattern width
Timer value range
2Eꢀ24
2Eꢀ24
2Eꢀ24
2Eꢀ24
32 bits
32 bits
128 bits
128 bits
50 ns to 2199 seconds
50 ns to 2199 seconds
Timer resolution
2 ns
2 ns
Timer accuracy
(5 ns ꢀ0.01ꢁ)
(5 ns ꢀ0.01ꢁ)
Greater than duration
Less than duration
3.33 ns to 55 ms in 3.3 ns increments
6.67 ns to 55 ms in 3.3 ns increments
50 ns
3.33 ns to 55 ms in 3.3 ns increments
6.67 ns to 55 ms in 3.3 ns increments
50 ns
Timer reset latency
16