Agilent 16950B and 16951B Specifications and Characteristics
State (Synchronous) Analysis Mode
300 Mb/s State Mode
667 Mb/s State Mode
Minimum state clock pulse width
Single edge
Multiple edge
1.0 ns
1.0 ns
500 ps
1.50 ns
Clock qualifier setup time
Clock qualifier hold time
Time tag resolution
500 ps
0
N/A
N/A
2 ns
1.5 ns
Maximum time count between stored states 32 days
32 days
Maximum trigger sequence speed
Maximum trigger sequence levels
Trigger sequence level branching
Trigger position
300 MHz
667 MHz
16
16
Arbitrary 4-way if/then/else
Start, center, end, or user-defined
2-way if/then/else
Start, center, end, or user-defined
Trigger resources
16 patterns evaluated as =, =/, >, ≥, <, ≤
14 double-bounded ranges evaluated as
in range, not in range
14 patterns evaluated as =, =/, >, ≥, <, ≤
7 double-bounded ranges evaluated as
in range, not in range
2 timers per module
2 global counters
1 occurrence counter per sequence level
4 flags
1 occurrence counter per sequence level
4 flags
Trigger resource conditions
Trigger actions
Arbitrary Boolean combinations
Arbitrary Boolean combinations
Go To
Go To
Trigger, send e-mail, and fill memory
Trigger and Go To
Trigger and fill memory
Store/don’t store sample
Turn on/off default storing
Timer start/stop/pause/resume
Global counter increment/decrement/reset
Occurrence counter reset
Flag set/clear
Store qualification
Maximum global counter
Maximum occurrence counter
Maximum pattern width
Maximum range width
Timers range
Default (global) and per sequence level
Default (global)
N/A
2Eꢀ24
2Eꢀ24
2Eꢀ24
128 bits
64 bits
N/A
128 bits
64 bits
50 ns to 2199 seconds
2 ns
Timer resolution
N/A
Timer accuracy
(5 ns ꢀ0.01ꢁ)
50 ns
N/A
Timer reset latency
14
N/A