欢迎访问ic37.com |
会员登录 免费注册
发布采购

16760A 参数 Datasheet PDF下载

16760A图片预览
型号: 16760A
PDF下载: 下载PDF文件 查看货源
内容描述: 逻辑分析仪 [Logic Analyzer]
分类和应用:
文件页数/大小: 35 页 / 1054 K
品牌: AGILENT [ AGILENT TECHNOLOGIES, LTD. ]
 浏览型号16760A的Datasheet PDF文件第11页浏览型号16760A的Datasheet PDF文件第12页浏览型号16760A的Datasheet PDF文件第13页浏览型号16760A的Datasheet PDF文件第14页浏览型号16760A的Datasheet PDF文件第16页浏览型号16760A的Datasheet PDF文件第17页浏览型号16760A的Datasheet PDF文件第18页浏览型号16760A的Datasheet PDF文件第19页  
Agilent 16950B and 16951B Specifications and Characteristics  
Timing (Asynchronous) Analysis Mode  
Sample rate on all channels  
Sample rate in half channel mode  
Number of channels  
Conventional Timing  
600 MHz  
Transitional Timing [9]  
600 MHz  
N/A  
1200 MHz  
68 x (number of modules)  
For sample rates < 600 MHz: 68 x (number  
of modules).  
For 600 MHz sample rate: 68 x (number of  
modules) – 34  
Maximum channels on a single time  
base and trigger  
340  
340  
Number of independent analyzers [6]  
Sample period (half channel)  
Minimum sample period (full channel)  
Minimum data pulse width  
2
2
833 ps  
N/A  
1.67 ns  
1.67 ns  
1 sample period ꢀ 500 ps  
1 sample period ꢀ 500 ps  
Time interval accuracy  
(1 sample period ꢀ 1.25 ns ꢀ 0.01ꢁ of  
time interval reading)  
(1 sample period ꢀ 1.25 ns ꢀ 0.01ꢁ of  
time interval reading)  
Memory depth in full channel mode –  
16950B  
Option 001: 1 M samples  
Option 004: 4 M samples  
Option 016: 16 M samples  
Option 032: 32 M samples  
Option 064: 64 M samples  
Option 001: 1 M samples  
Option 004: 4 M samples  
Option 016: 16 M samples  
Option 032: 32 M samples  
Option 064: 64 M samples  
Memory depth in full channel mode –  
16951B  
256 M samples  
256 M samples  
N/A  
Memory depth in half channel mode –  
16950B  
Option 001: 2 M samples  
Option 004: 8 M samples  
Option 016: 32 M samples  
Option 032: 64 M samples  
Option 064: 128 M samples  
Memory depth in half channel mode –  
16951B  
512 M samples  
N/A  
Maximum trigger sequence speed  
Maximum trigger sequence levels  
300 MHz  
16  
300 MHz  
16  
[6] Independent analyzers may be either state or timing. When the 600 Mb/s state mode is selected, only one analyzer may be used.  
[9] Transitional timing speed and memory depth are halved unless a spare pod pair (34 channels) is unassigned.  
15  
 复制成功!