14 | Keysight | N9923A FieldFox RF Vector Network Analyzer - Data Sheet  
General Information  
Description  
Specification  
Supplemental Information  
Calibration cycle  
1 year  
Environmental  
Keysight Technologies outdoor  
equipment class1  
MIL–PRF–28800F class 2  
Altitude – operating  
9,144 m (30,000 ft)  
Under battery operation  
[AC to DC adapter rated at 3,000 m (9,840 ft)]  
Altitude – non–operating  
Intrusion protection  
Temperature range  
Operating  
15,240 m (50,000 ft)  
IP 30 IEC/EN 60529  
AC power  
–10 to 55 °C  
–10 to 50 °C  
–51 to 71 °C  
Battery  
-10 to 55 °C (typical)  
Storage  
With the battery pack removed.  
The battery packs should be stored in an environment with low humidity.  
Extended exposure to temperature above 45 ºC could degrade battery  
performance and life.  
EMC  
Complies with European  
EMC  
Directive 2004/108/EC  
EC/EN 61326–1  
CISPR Pub 11 Group 1, class A  
AS/NZS CISPR 11  
ICES/NMB–001  
ESD  
IEC/EN 61000–4–2  
Safety  
Complies with European  
Low Voltage Directive  
2006/95/EC  
EC/EN 61010–1 2nd Edition  
Canada: CSA C22.2  
No. 61010–1–04  
USA: UL 61010–1 2nd Edition  
Power  
Power supply  
External DC input  
External AC power adapter  
Input  
15 to 19 VDC  
40 W maximum when battery charging  
Efficiency Level IV, 115 VAC  
100 to 250 VAC, 50 to 60 Hz  
1.25 – 0.56 A  
Output  
15 VDC, 4 A  
Power consumption  
On: 14 W (typical)  
1. Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual (ETM) for outdoor equipment (OE) and  