11 | Keysight | N9923A FieldFox RF Vector Network Analyzer - Data Sheet  
Mixed-Mode S-Parameters (Option 212)  
Mixed-mode S-parameters are also known as balanced measurements.  
Measurements  
Scc11  
Sdd11  
Scd11  
Sdc11  
Common mode reflection  
Differential mode reflection  
Differential mode stimulus, common mode response  
Common mode stimulus, differential mode response  
FieldFox’s mixed-mode S-parameter measurements require the use of the default factory calibration or a user 2-port calibration.  
So the FieldFox analyzer must be equipped with 2-port measurement functionality to measure mixed-mode S-parameters. Mixed-  
mode S-parameters are an extension of the VNA capabilities.  
Vector Voltmeter (Option 308)  
With vector voltmeter mode, you can characterize the difference between two measurements easily. The zeroing function allows you  
to create a reference signal, and characterize the difference between two device measurements.  
1-port cable trimming - reflection or S11 measurement, magnitude and phase  
– 2-port transmission - transmission or S21 measurement, magnitude and phase  
– A/B and B/A - ratio of two receivers or channels, magnitude and phase –Need an  
external signal generator for the A/B or B/A measurement (must order Option 122).  
The results are shown on a large display in digital format. Refer to the network analyzer  
section for magnitude and phase accuracy information.  
Cable and Antenna Analyzer (Option 305)  
The following CAT mode performance parameters are in addition to the VNA mode performance parameters specified above.  
Table1: Cable and antenna analyzer specifications  
Description  
Cable loss  
Specification  
Supplemental Information  
Display range  
Resolution  
0 to 100 dB  
0.01 dB  
Distance–to–Fault  
Range = [(number of points – 1) Number of points auto coupled according to start and stop distance entered  
Horizontal range  
/ frequency span * 2] * velocity  
factor * speed of light  
Horizontal  
Resolution  
Resolution = Range/ (number of Number of points settable by user  
points – 1)  
Bandpass mode  
Window types  
Maximum, medium, and minimum windows  
External USB Power Sensor Support (Option 302)  
The external USB power sensor option supports the Keysight Technologies, Inc. USB Power Sensors. List of supported sensors available  
from: http://www.keysight.com/find/usbsensorsforfieldfox.  