欢迎访问ic37.com |
会员登录 免费注册
发布采购

N9913A-211 参数 Datasheet PDF下载

N9913A-211图片预览
型号: N9913A-211
PDF下载: 下载PDF文件 查看货源
内容描述: [FieldFox Handheld Analyzers]
分类和应用:
文件页数/大小: 48 页 / 6709 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号N9913A-211的Datasheet PDF文件第15页浏览型号N9913A-211的Datasheet PDF文件第16页浏览型号N9913A-211的Datasheet PDF文件第17页浏览型号N9913A-211的Datasheet PDF文件第18页浏览型号N9913A-211的Datasheet PDF文件第20页浏览型号N9913A-211的Datasheet PDF文件第21页浏览型号N9913A-211的Datasheet PDF文件第22页浏览型号N9913A-211的Datasheet PDF文件第23页  
19 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet  
Mixed-Mode S-Parameters  
Mixed-mode S-parameters are also known as balanced measurements.  
Measurements  
Scc11  
Sdd11  
Scd11  
Sdc11  
Common mode reflection  
Differential mode reflection  
Differential mode stimulus, common mode response  
Common mode stimulus, differential mode response  
FieldFox’s mixed-mode S-parameter measurements require the use of the default factory calibration or a user 2-port calibration. So  
the FieldFox analyzer must be equipped with 2-port measurement functionality to measure mixed-mode S-parameters. Mixed-mode  
S-parameters are an extension of the VNA capabilities.  
Vector Voltmeter (VVM)  
With vector voltmeter mode, you can characterize the difference between two measurements easily. The zeroing function allows you to  
create a reference signal, and characterize the difference between two device measurements. The results are shown on a large display  
in digital format.  
Models  
Frequency range  
30 kHz to 4 GHz  
N991xA, N992xA  
N9913A  
N9914A  
30 kHz to 6.5 GHz  
30 kHz to 9 GHz  
N9915A, N9925A  
N9916A, N9926A  
N9917A, N9927A  
N9918A, N9928A  
N9950A  
30 kHz to 14 GHz  
30 kHz to 18 GHz  
30 kHz to 26.5 GHz  
300 kHz to 32 GHz  
300 kHz to 44 GHz  
300 kHz to 50 GHz  
N995xA  
N9951A  
N9952A  
Setup parameters  
1-port cable trimming  
2-port transmission  
A/B and B/A  
Reflection or S11 measurement, magnitude and phase  
Transmission or S21 measurement, magnitude and phase  
Ratio of two receivers or channels, magnitude and phase – Need an external signal generator for the A/B or B/A  
measurement  
Frequency (one CW frequency point), IF bandwidth - 10 Hz to 100 kHz, output power - Low or high  
Ratio accuracy (A/B and B/A)  
Must zero, before measuring DUT. Recommend using a high-quality power splitter or 6 dB attenuators to minimize uncertainty due to  
mismatch.  
Frequency  
Nominal  
± 1.0  
N991xA, N992xA, N995xA  
100 to 300 kHz1  
> 300 kHz to 1 MHz  
> 1 to 100 MHz  
> 100 to 300 MHz  
> 300 MHz to 1.5 GHz  
> 1.5 to 2 GHz  
± 0.4  
± 0.2  
± 0.4  
± 0.6  
± 1.0  
1. Does not apply to N995xA models, which start at 300 kHz.  
 复制成功!