15 | Keysight | FieldFox Handheld Analyzers 4/6.5/9/14/18/26.5/32/44/50 GHz - Data Sheet
Corrected Measurement Uncertainty for N9913A/14A/15A/16A/17A/18A and
N9925A/26A/27A/28A (continued)
Full 2-port calibration, 85052D 3.5 mm calibration kit, spec
Power level of -15 dBm, 10 Hz IF bandwidth, no averaging, battery saver off, and 60-minute warm-up time. Includes uncertainties due
to drift, noise, compression, and dynamic accuracy. Coverage factor of x1 applied to uncertainties, for ease of comparison with other
industry handheld analyzers.
Corrected performance (dB)
Directivity
≤ 0.5 GHz
42
0.5 to 8 GHz
38
8 to 20 GHz
36
20 to 26.5 GHz
30
Source match
37
31
28
25
Load match
38
33
29
24
Reflection tracking
Transmission tracking
± 0.005
± 0.070
± 0.006
± 0.135
± 0.009
± 0.320
± 0.012
± 0.500
Transmission uncertainty (S21, S12)
Reflection uncertainty (S11, S22)