08 | Keysight | M9485A PXIe Multiport Vector Network Analyzer 1 MHz/50 MHz to 9 GHz - Data Sheet
Test port output (source)
Test port output frequency
Description
Typical
Frequency range
Resolution
1 MHz to 9 GHz
1 Hz
Accuracy
See M9300A
Test port output power
Description
Typical
Nominal power (preset power)
0 dBm
Power range/Power sweep range
1 M to 6 GHz
6 G to 9 GHz
–90 dBm to max leveled power
–35 dBm to max leveled power
Max leveled power
Resolution
See M9376A
0.01 dB
Level accuracy1
Normal measurement
Leveling mode: Internal
1 M to 9 GHz, 0 dBm
Leveling mode: Open loop
1 M to 9 GHz, 0 dBm
± 0.5 dB
± 1.0 dB
Multi DUT measurement
Leveling mode: Internal/Open loop3
1 M to 9 GHz, 0 dBm
± 1.0 dB
Level linearity1, 2
Leveling mode: Internal
–20 dBm to max leveled power, 1 M to 6 GHz
–20 dBm to max leveled power, 6 G to 9 GHz
–85 dBm to –20 dBm, 1 M to 3 GHz
–85 dBm to –20 dBm, 3 G to 6 GHz
–30 dBm to –20 dBm, 6 G to 9 GHz
Leveling mode: Open loop
± 0.5 dB
± 1.0 dB
± 1.4 dB
± 1.5 dB
± 2.3 dB
–20 dBm to max leveled power, 1 M to 6 GHz
–20 dBm to max leveled power, 6 G to 9 GHz
–85 dBm to –20 dBm, 1 M to 3 GHz
–85 dBm to –20 dBm, 3 G to 6 GHz
–30 dBm to –20 dBm, 6 G to 9 GHz
± 0.5 dB
± 1.0 dB
± 2.7 dB
± 2.8 dB
± 3.6 dB
Test port output signal purity
Description
Typical
Harmonics (2nd or 3rd)
1 M to 9 GHz, 0 dBm
< –20 dBc
< –30 dBc
Non-harmonic spurious
1 M to 9 GHz, 0 dBm
1. System power calibration is required.
2. Level linearity given is relative to 0 dBm.
3. Multi DUT measurement is available only for the M9376A. When your configuration includes any of M9377A, the Multi DUT measurement is NOT available.