08 | Keysight | M9485A PXIe Multiport Vector Network Analyzer 1 MHz/50 MHz to 9 GHz - Data Sheet  
Test port output (source)  
Test port output frequency  
Description  
Typical  
Frequency range  
Resolution  
1 MHz to 9 GHz  
1 Hz  
Accuracy  
See M9300A  
Test port output power  
Description  
Typical  
Nominal power (preset power)  
0 dBm  
Power range/Power sweep range  
1 M to 6 GHz  
6 G to 9 GHz  
–90 dBm to max leveled power  
–35 dBm to max leveled power  
Max leveled power  
Resolution  
See M9376A  
0.01 dB  
Level accuracy1  
Normal measurement  
Leveling mode: Internal  
1 M to 9 GHz, 0 dBm  
Leveling mode: Open loop  
1 M to 9 GHz, 0 dBm  
± 0.5 dB  
± 1.0 dB  
Multi DUT measurement  
Leveling mode: Internal/Open loop3  
1 M to 9 GHz, 0 dBm  
± 1.0 dB  
Level linearity1, 2  
Leveling mode: Internal  
–20 dBm to max leveled power, 1 M to 6 GHz  
–20 dBm to max leveled power, 6 G to 9 GHz  
–85 dBm to –20 dBm, 1 M to 3 GHz  
–85 dBm to –20 dBm, 3 G to 6 GHz  
–30 dBm to –20 dBm, 6 G to 9 GHz  
Leveling mode: Open loop  
± 0.5 dB  
± 1.0 dB  
± 1.4 dB  
± 1.5 dB  
± 2.3 dB  
–20 dBm to max leveled power, 1 M to 6 GHz  
–20 dBm to max leveled power, 6 G to 9 GHz  
–85 dBm to –20 dBm, 1 M to 3 GHz  
–85 dBm to –20 dBm, 3 G to 6 GHz  
–30 dBm to –20 dBm, 6 G to 9 GHz  
± 0.5 dB  
± 1.0 dB  
± 2.7 dB  
± 2.8 dB  
± 3.6 dB  
Test port output signal purity  
Description  
Typical  
Harmonics (2nd or 3rd)  
1 M to 9 GHz, 0 dBm  
< –20 dBc  
< –30 dBc  
Non-harmonic spurious  
1 M to 9 GHz, 0 dBm  
1. System power calibration is required.  
2. Level linearity given is relative to 0 dBm.  