10 | Keysight | M9485A PXIe Multiport Vector Network Analyzer 1 MHz/50 MHz to 9 GHz - Data Sheet
M9485A PXIe Multiport Vector Network Analyzer Characteristics (Typical)
Option 2xx (Combination of M9377A and M9378A/B Configuration)
Corrected system performance
Description
Typical
System dynamic range at test port3
(Source: Standard Configuration, Receiver: Standard Configuration, Test Attenuator 0 dB IF Bandwidth = 10 Hz)
10 M to 50 MHz
50 M to 100 MHz
100 M to 200 MHz
200 M to 400 MHz
400 M to 800 MHz
800 M to 3 GHz
3 G to 5 GHz
65 dB
122 dB
128 dB
134 dB
140 dB
144 dB
141 dB
138 dB
135 dB
134 dB
5 G to 6 GHz
6 G to 7 GHz
7 G to 9 GHz
(Source: Standard Configuration, Receiver: Wide Dynamic Range Configuration, Test Attenuator 0 dB IF Bandwidth = 10 Hz,)
10 M to 50 MHz
50 M to 3 GHz
3 G to 5 GHz
5 G to 6 GHz
6 G to 7 GHz
7 G to 9 GHz
118 dB
160 dB
157 dB
154 dB
150 dB
149 dB
Extended Dynamic Range at direct receiver access input 2, 3
(Source: Standard Configuration, Receiver: Direct Access Configuration, Tin, Test Attenuator 0 dB, IF Bandwidth = 10 Hz)
10 M to 50 MHz
50 M to 3 GHz
3 G to 5 GHz
5 G to 6 GHz
6 G to 7 GHz
7 G to 9 GHz
119 dB
162 dB
159 dB
156 dB
153 dB
152 dB
1. Theꢀtestꢀportꢀsystemꢀdynamicꢀrangeꢀisꢀcalculatedꢀasꢀtheꢀdifferenceꢀbetweenꢀtheꢀtestꢀportꢀrmsꢀnoiseꢀfloorꢀandꢀtheꢀsourceꢀmaximumꢀoutputꢀpower.
Crosstalk does not limit the dynamic range.
2. Theꢀextendedꢀdynamicꢀrangeꢀisꢀcalculatedꢀasꢀtheꢀdifferenceꢀbetweenꢀtheꢀdirectꢀaccessꢀreceiverꢀinputꢀrmsꢀnoiseꢀfloorꢀandꢀtheꢀsourceꢀmaximumꢀoutput
power. Crosstalk does not limit the dynamic range. The extended dynamic range is only available in one-path transmission measurements.
3. The effective dynamic range must take measurement uncertainties and interfering signals into account.