10 | Keysight | M9485A PXIe Multiport Vector Network Analyzer 1 MHz/50 MHz to 9 GHz - Data Sheet  
M9485A PXIe Multiport Vector Network Analyzer Characteristics (Typical)  
Option 2xx (Combination of M9377A and M9378A/B Configuration)  
Corrected system performance  
Description  
Typical  
System dynamic range at test port3  
(Source: Standard Configuration, Receiver: Standard Configuration, Test Attenuator 0 dB IF Bandwidth = 10 Hz)  
10 M to 50 MHz  
50 M to 100 MHz  
100 M to 200 MHz  
200 M to 400 MHz  
400 M to 800 MHz  
800 M to 3 GHz  
3 G to 5 GHz  
65 dB  
122 dB  
128 dB  
134 dB  
140 dB  
144 dB  
141 dB  
138 dB  
135 dB  
134 dB  
5 G to 6 GHz  
6 G to 7 GHz  
7 G to 9 GHz  
(Source: Standard Configuration, Receiver: Wide Dynamic Range Configuration, Test Attenuator 0 dB IF Bandwidth = 10 Hz,)  
10 M to 50 MHz  
50 M to 3 GHz  
3 G to 5 GHz  
5 G to 6 GHz  
6 G to 7 GHz  
7 G to 9 GHz  
118 dB  
160 dB  
157 dB  
154 dB  
150 dB  
149 dB  
Extended Dynamic Range at direct receiver access input 2, 3  
(Source: Standard Configuration, Receiver: Direct Access Configuration, Tin, Test Attenuator 0 dB, IF Bandwidth = 10 Hz)  
10 M to 50 MHz  
50 M to 3 GHz  
3 G to 5 GHz  
5 G to 6 GHz  
6 G to 7 GHz  
7 G to 9 GHz  
119 dB  
162 dB  
159 dB  
156 dB  
153 dB  
152 dB  
1. Theꢀtestꢀportꢀsystemꢀdynamicꢀrangeꢀisꢀcalculatedꢀasꢀtheꢀdifferenceꢀbetweenꢀtheꢀtestꢀportꢀrmsꢀnoiseꢀfloorꢀandꢀtheꢀsourceꢀmaximumꢀoutputꢀpower.  
Crosstalk does not limit the dynamic range.  
2. Theꢀextendedꢀdynamicꢀrangeꢀisꢀcalculatedꢀasꢀtheꢀdifferenceꢀbetweenꢀtheꢀdirectꢀaccessꢀreceiverꢀinputꢀrmsꢀnoiseꢀfloorꢀandꢀtheꢀsourceꢀmaximumꢀoutput  
power. Crosstalk does not limit the dynamic range. The extended dynamic range is only available in one-path transmission measurements.  
3. The effective dynamic range must take measurement uncertainties and interfering signals into account.  
 复制成功!