12 | Keysight | M9485A PXIe Multiport Vector Network Analyzer 1 MHz/50 MHz to 9 GHz - Data Sheet
Corrected system performance with calibration kit
Corrected system performance with 3.5 mm device connector type, 85033E
calibration kit
Network analyzer: M9485A-2xx (Combination of M9377A/M9378A/B)
Calibrationꢀkit:ꢀ
85033Eꢀ(3.5ꢀmm,ꢀ50ꢀΩ)
Calibration:
full 2-port
IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23
°C (±3 °C) with < 1 °C deviation from calibration temperature, isolation calibration
performed
Typical (dB)
Description
50 M to 400 MHz
400 M to 800 MHz
800 M to 3 GHz
3 G to 9 GHz
38
Directivity
46
46
44
Source match
Load match
43
43
40
36
46
46
44
38
reflection tracking
Transmission tracking
± 0.006
± 0.034
± 0.006
± 0.034
± 0.007
± 0.045
± 0.010
± 0.079
Transmission uncertainty (typical)
Reflection uncertainty (typical)