23 | Keysight | InfiniiVision 1000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
WaveGen – Built-in function waveform generator (specifications are typical)
Note: Only available on WaveGen models EDUX1002G and DSOX1102G� WaveGen cannot be added to the other models�
EDUX1002G/DSOX1102G
WaveGen out
Waveforms
Modulation
Front-panel BNC connector
Sine, square, ramp, pulse, DC, noise
Modulation types: AM, FM, FSK
Carrier waveforms: Sine, ramp
Modulation source: Internal (no external modulation capability)
AM:
– Modulation: sine, square, ramp
– Modulation frequency: 1 Hz to 20 kHz
– Depth: 0 to 100%
FM:
– Modulation: Sine, square, ramp
– Modulation frequency: 1 Hz to 20 kHz
– Minimum carrier frequency: 10 Hz
– Deviation: 1 Hz to carrier frequency or (2e12 / carrier frequency), whichever is smaller
FSK:
– Modulation: 50% duty cycle square wave
– FSK rate: 1 Hz to 20 kHz
– Hop frequency: 2 x FSK rate to 10 MHz
Frequency range: 0.1 Hz to 20 MHz
Amplitude flatness: ± 0.5 dB (relative to 1 kHz)
Harmonic distortion: —40 dBc
Sine
Spurious (non harmonics): —40 dBc
Total harmonic distortion: 1%
SNR (50 Ω load, 500 MHz bandwidth): 40 dB (typical); 30 dB (min)
Frequency range: 0.1 Hz to 10 MHz
Duty cycle: 20 to 80%
Square wave /pulse
Duty cycle resolution: Larger of 1% or 1 ns (≥ 25 kHz) or 10 ns (< 25 kHz)
Pulse width: 20 ns minimum
Rise/fall time: 18 ns (10 to 90%)
Pulse width resolution: 1 ns (≥ 25 kHz) or 10 ns (< 25 kHz) or 5 digits, whichever is larger
Overshoot: < 2%
Asymmetry (at 50% DC): ± 1% ± 5 ns
Jitter (TIE RMS): 500 ps
Ramp/triangle wave
Noise
Frequency range: 0.1 Hz to 200 kHz
Linearity: 1%
Variable symmetry: 0 to 100%
Symmetry resolution: 1%
Bandwidth: 20 MHz typical