14 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet  
MFCMU (multi frequency capacitance measurement unit) module specifications  
Measurement functions  
DC bias function  
Measurement accuracy  
The following parameters are used to  
express the impedance measurement  
accuracy at the measurement port of the  
MFCMU or the MFCMU cable  
(1.5 m or 3 m).  
Measurement parameters:  
Cp-G, Cp-D, Cp-Q, Cp-Rp, Cs-Rs, Cs-D,  
Cs-Q, Lp-G, Lp-D, Lp-Q, Lp-Rp, Ls-Rs,  
Ls-D, Ls-Q, R-X, G-B, Z-θ, Y-θ  
DC bias:  
Range: 0 to ±25 V  
Resolution: 1 mV  
Accuracy: ±(0.5% + 5.0 mV)  
at the measurement port or the  
MFCMU or the MFCMU cable (1.5  
m/3 m)  
Ranging:  
Auto and fixed  
ZX: Impedance measurement value (Ω)  
DX: Measurement value of D  
E = EP’ + (ZS’/|ZX| + YO’|ZX|) x 100 (%)  
EP’ = EPL + EPOSC + EP (%)  
Measurement terminal:  
Four-terminal pair configuration, four  
BNC (female) connectors  
Maximum DC bias current  
(Supplemental characteristics):  
Cable length:  
1.5 m or 3 m, automatic  
identification of accessories  
Impedance  
measurement range  
Maximum DC  
bias current  
YO’ = YOL + YOSC + YO (S)  
50 Ω  
10 mA  
10 mA  
10 mA  
1 mA  
ZS’ = ZSL + ZOSC + ZS (Ω)  
100 Ω  
300 Ω  
1 kΩ  
Test signal  
|Z| accuracy  
±E (%)  
Frequency:  
3 kΩ  
1 mA  
Range: 1 kHz to 5 MHz  
Resolution: 1 mHz (minimum)  
Accuracy: ±0.008%  
10 kΩ  
30 kΩ  
100 kΩ  
300 kΩ  
100 µA  
100 µA  
10 µA  
10 µA  
θ accuracy  
±E/100 (rad)  
C accuracy  
at DX ≤ 0.1  
±E (%)  
Output signal level:  
Range: 10 mVrms to 250 mVrms  
Resolution: 1 mVrms  
Output impedance: 50 Ω, typical  
DC bias monitor:  
Range: 0 to ±25 V  
Accuracy:  
at DX > 0.1  
±(10.0% + 1 mVrms) at the  
measurement port of the MFCMU  
±(15.0% + 1 mVrms) at the  
measurement port of MFCMU cable  
(1.5 m or 3 m)  
Accuracy (open load):  
2
±E x (1+DX )(%)  
±(0.2% of reading + 10.0 mV)  
at the measurement port or the MFCMU  
cable (1.5 m/3 m)  
D accuracy  
at DX ≤ 0.1  
±E/100  
Output impedance: 50 Ω, typical  
Signal level monitor:  
Range: 10 mVrms to 250 mVrms  
Accuracy:  
Sweep characteristics  
at DX > 0.1  
±E x (1 + DX)/100  
Available sweep parameters:  
Oscillator level, DC bias voltage,  
frequency  
±(10.0% of reading + 1 mVrms) at the  
measurement port of the MFCMU  
G accuracy  
at DX ≤ 0.1  
±E/ DX (%)  
Sweep type: linear, log  
±(15.0% + 1 mVrms  
)
Sweep mode: single, double  
Sweep direction: up, down  
Number of measurement points:  
Maximum 1001 points  
at the measurement port of MFCMU  
cable (1.5 m or 3 m)  
at DX > 0.1  
2
±E x (1+DX )/DX (%)  
Note: measurement accuracy is specified  