11 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet
MCSMU Module Specifications
Voltage range, resolution, and accuracy
Voltage
range
Force
resolution
Measure
resolution
Force accuracy 1
±(% + mV)
Measure accuracy 1
(% + mV)
Maximum
current
±0.2 V
±2 V
±20 V
±40 V2
200 nV
2 µV
20 µV
200 nV
2 µV
20 µV
±(0.06 + 0.14)
±(0.06 + 0.6)
±(0.06 + 3)
±(0.06 + 0.14)
±(0.06 + 0.6)
±(0.06 + 3)
1 A
1 A
1 A
1 A
40 µV
40 µV
±(0.06 + 3)
±(0.06 + 3)
1. ±(% of reading value + fixed offset in mV).
2. Maximum output voltage is 30 V.
Current range, resolution, and accuracy
Current
range
Force
resolution
Measure
resolution
Force accuracy 1
(% + A + A)
Measure accuracy 1
(% + A + A)
Maximum
voltage
±10 µA
±100 µA
±1 mA
±10 mA
±100 mA
±1 A2
10 pA
100 pA
1 nA
10 nA
100 nA
1 µA
10 pA
100 pA
1 nA
10 nA
100 nA
1 µA
±(0.06 + 2E-9 + Vo x 1E-10)
±(0.06 + 2E-8 + Vo x 1E-9)
±(0.06 + 2E-7 + Vo x 1E-8)
±(0.06 + 2E-6 + Vo x 1E-7)
±(0.06 + 2E-5 + Vo x 1E-6)
±(0.4 + 2E-4 + Vo x 1E-5)
±(0.06 + 2E-9 + Vo x 1E-10)
±(0.06 + 2E-8 + Vo x 1E-9)
±(0.06 + 2E-7 + Vo x 1E-8)
±(0.06 + 2E-6 + Vo x 1E-7)
±(0.06 + 2E-5 + Vo x 1E-6)
±(0.4 + 2E-4 + Vo x 1E-5)
30 V
30 V
30 V
30 V
30 V
30 V
1. ±(% of reading value + fixed offset in A + proportional offset in A), Vo is the output voltage in V.
2. Pulse mode only. The maximum value of the base current during pulsing is ±50 mA.
Power consumption
MCSMU measurement and output range
Pulse only
Pulse only
Voltage source mode:
Voltage range
Power
DC and pulse
DC and pulse
Current (A)
1
0.2 V
2 V
40 V
40 x Ic (W)
40 x Ic (W)
40 x Ic (W)
Where Ic is the current compliance setting.
0.1
Voltage (V)
30
Current source mode:
-30
Voltage
Power
-0.1
compliance
Vc ≤ 0.2
0.2 < Vc ≤ 2
2 < Vc ≤ 40
40 x Io (W)
40 x Io (W)
40 x Io (W)
-1
Where Vc is the voltage compliance setting and
Io is output current.