BLOCK CHARACTERISTICS
GStage Block A10858
The stage block is for using an infinity-corrected objective lens in object
image generation and fluorescence detection. It has two C-mount ports on
the top and bottom of its rear panel. You can use other optical blocks to
connect the stage block to a PMT module or ICCD camera (p. 27). Because
you can use the stage block cover to keep the measured area in darkness,
the stage block is useful even when detecting faint light. Furthermore, if you
remove the cover and illuminate the sample appropriately using an
illuminator, you can generate images by using the stage block with the
A10859-01 tube lens block, CCD cameras, and other devices.
SPECIFICATIONS
Travels
Compatible
Type No.
Weight
Port
X-Axis
Z-Axis
Objective Lens
Infinity-corrected Objective lens
M20.32 × P0.706
Parfocal distance*: 45 mm
Approx.
4 kg
A10858
6.5 mm
3 mm 2; C-mount
* For information about the parfocal distance, see p. 14.
Dimensional Outlines (Unit: mm)
110
50
(38)
160
2-C-MOUNT
Directions that you can
move the workplate in
using a micrometer.
Direction of beam
movement caused
by beam steering
mirror assembly
adjustment
TPMOA0049EA
Workplate and Beam Movement
From outside of the stage block, you can use a micrometer to move the
workplate horizontally along the X axis or vertically along the Z axis.
You can also use the three setup screws in the beam steering mirror
assembly to change the internal mirror angle and adjust the horizontal
Y axis.
Beam steering
mirror assembly
setup screws
Accesories
Dimensional Outlines (Unit: mm)
Stage Block Rack A11058
4-M6
204
250
Sizes of the included spacers
Spacer 1
Spacer 2
Spacer 3
90 mm × 90 mm, Thickness: 10 mm
60 mm × 60 mm, Thickness: 18 mm
60 mm × 60 mm, Thickness: 25 mm
2
1
1
HEIGHT-ADJUSTMENT BOLT
The spacer sizes have been designed according to their uses.
TPMOA0066EA
12